For full functionality of this site it is necessary to enable JavaScript.

HT HT10 Two pole multifucntion tester (12~1000V; 0 ~ 1999Ω)

DC/AC VOLTAGE MEASUREMENT (LED indications)

Range [V]; ±12, 24, 50, 120, 230, 400, 690, 1000

Resolution [V]: 1

Frequency range [Hz]: 0/40Hz ~ 400Hz

Accuracy: Compliance with IEC/EN61243- 3:2014

Overload protection: 1000VDC/AC 


DC/AC VOLTAGE MEASUREMENT (Display indications) 

Range [V]: 6 : 1000

Resolution [V]: 1

Frequency range [Hz]: 0/40Hz ~ 400Hz ±(3%rdg + 5dgt)

Accuracy:±(3%rdg + 5dgt) 

Overload protection: 1000VAC/DC 


AC VOLTAGE MEASUREMENT WITH LOW IMPEDANCE (LoZ) 

Range [V]: 6 ~ 1000 

Resolution [V]: 

Frequency range:  40Hz ~ 400Hz 

Input impedance: ca 7k

Overload protection : 1000VAC/DC 


AC VOLTAGE WITH 1 WIRE

Range [V]: 100 ~ 1000 

Frequency range: 50 ~ 400 

Measurement method: 1-wire with contact on metal live parts 

Overload protection: 1000VAC 


RESISTANCE MEASUREMENT 

Range []: 0  1999 

Resolution []: 1 

Accuracy: ±(5.0%rdg+10dgt)  

Test current:  <30µA 

Overload protection: 1000VAC/DC 


CONTINUITY TEST 

GENERAL SPECIFICATIONS

Dimensions (L x W x H): 270 x 70 x 30mm

Weight (included batteries): 290g 

Internal supply: 2x1.5V alkaline batteries type AAA, IEC LR03

Auto Power OFF: after 1minutes of idleness (resistance measurement) 


Standard accessories

Carrying bag

Batteries and Manual

Details

Datasheet


Manual

  • 质量承诺
  • 正品保修
  • 送货到家
  • 交易简单化