JEIOTECH TH-ICH-800 温度和湿度控制箱 (760L,230V/50Hz)
制造商: JEIOTECH Model: TH-ICH-800 P/N: AAHC1012K - 报价要求
- 联系
Chamber Volume (L / cu ft) 760 lit
Controller PID control (LCD Type)
Temperature: -5 to 85 / 23 to 185 without Light
Fluctuation 1) (±℃ / oF) at 40℃ / 60%RH 0.3 / 0.55
Variation 1) (±℃ / oF) at 40℃ / 60%RH 0.5 / 0.9
Humidity: 20 to 85 %
:Fluctuation (±%RH) at 60%RH / 40℃ 3
: Variation 2) (±%RH) at 60%RH / 40℃ 5
Light Source (ICH Q1B option 2): Cool White, Ultraviolet-A, Light Uniformity
Time: 25min (20℃ → 85℃)/ 35min (20℃ → -5℃)
Refrigeration: System Air - cooled, R-404A, Capacity (W) 470
Material: Ni-Cr Wire Heater / SUS Tube Heater , Polyurethane
Interface Port Basically RS-232C, [RS-485 : optional]
Dimension (W×D×H): Internal (mm / inch) 750×880×1160mm
: External (mm / inch) 1220×1215×1965mm
Net Weight (kg / lbs) 400±10
Electrical Requirements AC230V, 1ph, 50Hz 23.5 A Cat.No. AAHC1012K
Temperature & Humidity Control range
- TH-TG (190 to 960L)
- 40~90% RH at 10℃
- 20~95% RH at 30℃
- 10~95% RH at 50~80℃
- 10~80% RH at 90℃
- 40~90% RH at 15℃
- 30~95% RH at 30℃
- 20~95% RH at 50~80℃
- 20~85% RH at 85℃
- 40~80% RH at 15℃
- 30~85% RH at 30℃
- 20~85% RH at 50~85℃
Constructional Features
- High quality #304 stainless steel exterior and interior.
- Casters for easy mobility when installing or moving environmental test chamber.
- Convenient maintenance
- Available to check total operating time. - Unique door lock & trim.
- Vice lock type door and foam silicone packing to ensure a perfect seal.
- Cable port for external probes or wires.
- 50mm diameter cable port comes standard on the left side of the unit. Additional 50mm or 80mm diameter cable ports can be custom ordered. (optional)
- Container type tank for easy adding and changing of water.
- Maintenance of the water tank can be easily performed from the front side. - Door lock.
- Cyclic defrost functions does not affect the temperature and humidity control for extended operation.
(Under 15℃ set value) - Powerful two Sirocco fans maintains a constant airflow in the chamber.
Safety Features
- Leakage breaker for power supply.
- Over current protection.
- Overheat / overcool protection.
- Switch-off after alarm for over heating.
- Water empty alarm.
- Door open alarm.
- Vice lock door type locking mechanism.
Operating Features
- Microprocess PID control / Auto-tuning / Calibration.
- Digital LCD display with touch-sensitive 7 keypads.
- Easy operation with wide LCD screen.
- Temperature and humidity / timer settings and operation condition. - Program patterns are memorized for 10 years due to a battery built into the PCB.
- Expansion parameter inputs. (conveniently display by letters)
- Programming operation.
- Computer interface.
- Software provided.
- Save data in excel format.
- Storage of program information, backup of value settings / recovery, and temperature / humidity values.
- RS-232C port as standard offer. - Chart recorder. (optional)
- 6-point dots type recorder with digital display.
- Digital recorder. (6-channels) - In the event of a validation the apparatus correctly controls each light and then transfers accurate data and is equipped with light sensors. (UV light sensor / VIS light sensor)
- Samples that have to be exposed to the light in which it is completely near the UV energy of less than 200W-hr/m2 is provided to compare directly the results from overall intensity of less than 1.20 million lux-hr, drug substances and reagent.
- Total fluorescent lamp and UV lamp indicator and operating time indicator.
Safety Features
- Leakage breaker for power supply.
- Over current protection.
- Overheat protection.
- Switch-off after alarm for over heating.
- Door open alarm.
- Water empty.
- Vice lock door type locking mechanism. (optional)
- Alarm for completed Photostability Test.
Model | TH-ICH-300 | TH-ICH-800 | ||
---|---|---|---|---|
Chamber Volume (L / cu ft) | 300 / 10.6 | 760 / 26.8 | ||
Controller | PID control (LCD Type) | |||
Temperature 1) | Range | without Humidity (℃ /oF) | -5 to 85 / 23 to 185 without Light | |
0 to 85 / 32 to 185 with Light | ||||
with Humidity (℃ / oF) | +15 to 85 / 59 to185 with Light | |||
Fluctuation 1) (±℃ / oF) at 40℃ / 60%RH | 0.3 / 0.55 | 0.3 / 0.55 | ||
Variation 1) (±℃ / oF) at 40℃ / 60%RH | 0.5 / 0.9 | 0.7 / 1.3 | ||
Humidity | Range (%RH) | FULL RANGE : 20 to 85 40 to 80 at 15℃ (59oF) 30 to 85 at 30℃ (86oF) 20 to 85 at 50℃ to 85℃ (122oF to 185oF) |
||
Fluctuation 2) (±%RH) at 60%RH / 40℃ | 3 | |||
Variation 2) (±%RH) at 60%RH / 40℃ | 5 | |||
Light Source (ICH Q1B option 2) |
Cool White 3) | 6,000 lux | 6,500 lux | |
Ultraviolet-A 3) | 4.5 W/㎡ | 5 W/㎡ | ||
Light Uniformity | VIS. : ±10% UVA : ±15% | VIS. : ±15% UVA : ±15% | ||
Typical Time Required to Reach ICH Recommendation of 1.2 million lux-hr | ≒ 200Hr | ≒ 184Hr | ||
Typical Time Required to Reach ICH Recommendation of 200 W-hr/m2 | ≒ 50Hr | ≒ 40Hr | ||
Time | Heating 4) | 25min (20℃ → 85℃) | 30min (20℃ → 85℃) | |
Cooling 4) | 35min (20℃ → -5℃) | 40min (20℃ → -5℃) | ||
Refrigeration | System | Air - cooled | ||
Refrigerant | R- 404A | |||
Capacity (W) | 250 | 470 | ||
Cooler | Copper plate fin cooler | |||
Heater | Material | Ni-Cr Wire Heater / SUS Tube Heater | ||
Drying Bulb (W) | 1500 | 2300 | ||
Humidity Bulb (W) | 1200 | 2000 | ||
Sensor | Temperature | Pt 100 | ||
Humidity | Electronic sensor (Free of drying a wet pack and good response time) | |||
Air Flow | 2 set of sirocco fan (2×60 W) : Vertical type | |||
Material | Internal | SUS #304 (2B polish) | ||
External | SUS #304 (Hairline finish) & Steel, 2t, Double painted & baked | |||
Insulation | Polyurethane (70mm / 2.76") | |||
Door | EPDM Sheet | |||
Operating Temperature | 18℃ to 30℃ / 64.4oF to 56oF, 85%RH (Non-corrosive and pure place) | |||
Water Tank | Capacity (L / cu ft) | 15 / 0.53 | ||
Water Quality | pH6.2 ~ 7.2, Electrical conductivity 20㎲ / cm to below | |||
Interface Port | Basically RS-232C, [RS-485 : optional] | |||
Cable Port | Basically one hole provided / Ø 50mm (2"),Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2") | |||
Dimension (W×D×H) |
Internal (mm / inch) | 750×650×650 / 29.5×25.6×25.6 | 750×880×1160 / 29.5×34.6×45.7 | |
External (mm / inch) | 1130×985×1520 / 44.5×38.8×59.8 | 1220×1215×1965 / 48×47.8×77.4 | ||
Net Weight (kg / lbs) | 280±10 / 617±22 | 400±10 / 882±22 | ||
Electrical Requirements | AC230V, 1ph, 60Hz | 15.4 A | 24.5 A | |
Cat.No. | AAHC1001K | AAHC1011K | ||
AC230V, 1ph, 50Hz | 14.7 A | 23.5 A | ||
Cat.No. | AAHC1002K | AAHC1012K |
2) Technical data according to DIN 12880, IEC 60068.
3) The value is measured on the center of chamber at the point of vertical direction.
4) Up to 98% of the set value.