- Detector: gamma CsI(Tl)
- Sensitivity:
+ for 137Сs, no less 100 (s-1)/(μSv/h) (1.0 (s-1)/(μR/h))
+ for 241Am, no less 100 (s-1)/(μSv/h) (1.0 (s-1)/(μR/h))
- Energy range: gamma 0.033 – 3.0 MeV
- Dose Rate: gamma 0.01 – 99.99 μSv/h (1 – 9999 μR/h)
- Accuracy: ±30% (in range 0.1 – 70 μSv/h (10 – 7000 μR/h) )
- Response time: 0.25 s
- Radionuclide identification using bluetooth communication with external Pocket PC or smartphone :
+ Special nuclear materials (SNM): X 233U, 235U, 237Np, Pu
+ Medical radionuclides: 18F, 67Ga, 51Cr, 75Se, 89Sr, 99Mo, 99mTc,103Pd, 111In, 123I, 131I,153Sm, 201Ti, 133Xe
+ Naturally occurring radioactive materials (NORM): 40K, 226Ra, 232Th and daughters, 238U and daughters
+ Industrial radionuclides: 57Co, 60Co, 133Ba, 137Cs, 192Ir,
226Ra, 241Am
- Standards compliance: ITRAP/IAEA requirements, ANSI N42.32, ANSI N42.33(1), IEC 62401
- Alarm type: visual, audio, vibration
- Data recording: 1000
- Environmental protection: IP65
- Drop test on concrete floor: 1.5 m (4.9 ft)
- Power supply: one AA battery
- Battery life time: up to 1000 hours
- Operating temperature: -30°C to 50°C (-22°F to 122°F)
- Size (without cover): 103 x 57 x 32 mm (4 x 2¼ x 1¼)
- Weight (without cover): 280 g (9.9 oz)
- Low battery warning: LCD
- Overload indication: gamma OL
- PC Communication: IRDA