- Detector:
+ gamma CsI(Tl), GM counter
+ Neutron LiI (Eu)
- Sensitivity:
+ for 137Сs, no less 85 (s-1)/(μSv/h) (0.85 (s-1)/(μR/h))
+ for 241Am, no less 100 (s-1)/(μSv/h) (1 (s-1)/(μR/h))
- Energy range :
+ gamma 0.033 - 3.0 MeV
+ Neutron from thermal to 14.0 MeV
- Dose Rate:
+ gamma 0.01 μSv/h – 10 Sv/h (1μR/h – 1000 R/h)
+ Neutron 0.1 - 999 s-1
- Dose: 0.001 mSv – 10 Sv (0.1 mR – 1000 R)
- Accuracy: ±(20+K1/H+K2*H )% where H - DER value in mSv/h; К1 – coefficient 0.0025 mSv/h, К2 – coefficient 0.002 (mSv/h)-1
- Response time: X
- Radionuclide identification using Bluetooth communication with external Pocket PC or smartphone:
+ Special nuclear materials (SNM): X
+ Medical radionuclides : X
+ Naturally occurring radioactive materials (NORM): X
+ Industrial radionuclides: X
- Standards compliance: X
- Alarm type: Audible, vibration, visible
- Data recording: up to 1000 data points
- Environmental protection: IP65
- Drop test on concrete floor: 0.7 m (2.3 ft) without cover 1.5 m (4.9 ft) with cover
- Power supply: 1.5 V (one AA (LR6) battery)
- Battery life time: not less than 1000 h
- Operating temperature: from -30oC to +50oC
- Size (without cover): 75 x 35 x 98 mm
- Weight (without cover): not more than 250 g
- Low battery warning: indication on LCD
- PC Communication: USB