For full functionality of this site it is necessary to enable JavaScript.

TTI MX100TP 三路输出可编程直流电源 (3Channel, 0~70V, 0~6A, 315W, USB/RS232/LAN)

Output 1
Ranges: Range 1 - 0V to 35V at 1mA to 3A; Range 2 - 0V to 16V at 1mA to 6A
Operating Mode: Constant voltage or constant current with automatic cross-over
and mode indication.
Voltage Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 1mV.
Current Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 1mA.
Setting Accuracy: Voltage - 0.05% of reading ± 3mV;
Current - 0.3% of reading ± 3mA to 3A, 0.5% of reading ± 3mA to 6A
Setting Stores: Up to 50 set-ups can be saved and recalled via the keyboard
(or the digital interfaces on MX100TP).
Load regulation: <0.01% +5mV (CV mode) for any load change using remote sense.
Line regulation: <0.01% +5mV (CV mode) for a 10% line voltage change.
Ripple & Noise: Typically <0.5mv rms="" 5mv="" pk-pk="" 1mv="" rms="" max="" 20mhz="" bandwidth="" br=""> Transient Response: <100us to="" within="" 50mv="" of="" set="" level="" for="" 5="" to="" 95="" load="" change="" br=""> Over Voltage Trip: Settable 1V to 40V in 0.1V steps
Over Current Trip: Settable 0.1A to 7A in 0.01A steps
Sensing: Selectable local or remote sensing.
Output 2
Ranges: Range 1 - 0V to 35V at 1mA to 3A; Range 2 - 0V to 16V at 1mA to 6A
Range 3 - 0V to 35V at 1mA to 6A* (available when output 3 is disabled).
Operating Mode: Constant voltage or constant current with automatic cross-over
and mode indication.
Voltage Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 10mV.
Current Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 10mA.
Setting Accuracy: Voltage - 0.1% of reading ± 10mV;
Current - 0.3% of reading ± 3mA to 3A, 0.5% of reading ± 3mA to 6A
Setting Stores: Up to 50 set-ups can be saved and recalled via the keyboard
(or the digital interfaces on MX100TP).
Load regulation: <0.01% +5mV (CV mode) for any load change using remote sense.
Line regulation: <0.01% +5mV (CV mode) for a 10% line voltage change.
Ripple & Noise: Typically <0.5mv rms="" 5mv="" pk-pk="" 1mv="" rms="" max="" 20mhz="" bandwidth="" br=""> Transient Response: <100us to="" within="" 50mv="" of="" set="" level="" for="" 5="" to="" 95="" load="" change="" br=""> Over Voltage Trip: Settable 1V to 40V in 0.1V steps
Over Current Trip: Settable 0.1A to 7A in 0.01A steps
Sensing: Selectable local or remote sensing.
Output 3
Ranges: Range 1 - 0V to 35V at 1mA to 3A; Range 2 - 0V to 70V at 1mA to 1.5A
Range 3 - 0V to 70V at 1mA to 3A* (available when output 2 is disabled).
Operating Mode: Constant voltage or constant current with automatic cross-over
and mode indication.
Voltage Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 10mV.
Current Setting: By direct numeric entry or quasi-analog rotary wheel; resolution 1mA.
Setting Accuracy: Voltage - 0.1% of reading ± 10mV;
Current - 0.3% of reading ± 3mA
Setting Stores: Up to 50 set-ups can be saved and recalled via the keyboard
(or the digital interfaces on MX100TP).
Load regulation: <0.01% +5mV (CV mode) for any load change using remote sense.
Line regulation: <0.01% +5mV (CV mode) for a 10% line voltage change.
Ripple & Noise: Typically <0.5mv rms="" 5mv="" pk-pk="" 1mv="" rms="" max="" 20mhz="" bandwidth="" br=""> 70V range - typically <1mv rms="" 10mv="" pk-pk="" 1="" 5mv="" rms="" max="" br=""> Transient Response: <100us to="" within="" 50mv="" of="" set="" level="" for="" 5="" to="" 95="" load="" change="" br=""> Over Voltage Trip: Settable 1V to 80V in 0.1V steps
Over Current Trip: Settable 0.1A to 3.5A in 0.01A steps
Sensing: Selectable local or remote sensing
Input
AC Input: 110V to 240V AC ±10%, 50/60Hz. Installation Category II.
Input Power: 500VA max.
Temperature & Environmental
Operating Range: +5ºC to +40ºC, 20% to 80% RH

GPIB interfaces: no, this feature has been discontinued

Details