Generalities
Basket capacity: 130 lt
Basket useful dimensions: 614x500x425 (WxDxH) mm approx
Maximum basket load: 50 Kg
Chamber external dimensions: 1252x2072x2844 (WxDxH) mm approx
Rated power: 15,1 kW
Rated current absorption: 28,1 A
Sound pressure level 60 dB(A) measured at 1m distance from the front of the unit, in 1,6m height, free field according to EN ISO 11201
Voltage: 400 V ±10%/50Hz/3ph + N + G
Weight: 1310 Kg
Performances (measured at room temperature of 22°C, empty working space and nominal voltage)
Temperature range:
upper chamber +70/+220°C
lower chamber -80/+100°C
Basket displacement speed <10 sec.
Temperature fluctuation: ±1°C
Temperature Rate of Change (measured at the control probe, according to IEC 60068-3-5):
Cold chamber:
heating: 2.0°C/min. from -80°C to +100°C
cooling: 4.7°C/min. from +100°C to -80°C
Hot chamber:
heating: 6.6°C/min. from +20°C to +220°C
Max load with MIL 883 test (15 min resetting time on specimen):
13 kg Ics distributed on 2 shelves with MIL 883 test D
20 kg Ics distributed on 2 shelves with MIL 883 test C
26 kg Ics distributed on 2 shelves with MIL 883 test B
Water consumption: 2,8 m3/h at a cooling water temperature of +29°C and temperature difference of 5°C (water temperature +12°C to +29°C)