Range Measuring range Measuring current Accuracy [Slow]
100mΩ 0.00mΩ~150.00mΩ 100mA within ±0.02%rdg±10μΩ
±3digit [Average]
±4digit [Slow]
±5digit [Fast]
1Ω 0.0000Ω~ 1.5000Ω 10mA refer to above
10Ω 0.000Ω~ 15.000Ω 1mA refer to above
100Ω 0.00Ω~ 150.00Ω 1mA refer to above
1kΩ 0.0Ω~ 1500.0Ω 1mA refer to above
10kΩ 0Ω~ 15000Ω 0.1mA refer to above
% 1mΩ~9.99mΩ/±50.0% Refer to the above within [(Range/Standard) /5]×0.01%±βdigit
10mΩ~10kΩ/±50.00% Refer to the above within [(Range/Standard) /5] ×0.01%±αdigit
Open-circuit voltage of measuring terminal: about 8V
Measuring method: 4-terminal measurement (with contact check)
Sampling time:
[Free running mode] 2~9 times per second (standard setting)
[Remote start mode] about 18.6msec./31.1msec. (standard setting)
Comparator set range:
[mΩ measurement] 0~15000 both the upper and lower limit.
[% measurement] ±50.00% both the upper and lower limit. (Standard 100~999:±50.0%)
Indication of comparator's comparison result: LED indication LO/GO/HI and buzzer
Control signal:
Remote start input:"L" [0V] →"H" [DC12V] start
Trimming signal:Open and "H" [DC12V]:Free run/"L" [0V] :Hold
Comparison output [LO/GO/HI]:Open-collector output max.40V, 100mA
Contact error output [CE]:Open-collector output max.40V, 100mA
End of comparison output [EOC]:Open-collector output max.40V, 100mA
Operation condition: [Temp.] +5℃~+40℃ [Humidity] less than 85%
Power supply: AC100V~240V selectable, 50/60Hz, about 60VA
Outer dimension: about 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.)
Weight: about 3.8kg