Range Measuring range Resolution Measuring current Accuracy
20mΩ 0.00mΩ~19.99mΩ 10μΩ 1A within ±0.2%rdg±2digit
200mΩ 0.0mΩ~199.9mΩ 100μΩ 1A within ±0.2%rdg±2digit
2000mΩ 0mΩ~ 1999mΩ 1mΩ 100mA within ±0.2%rdg±2digit
20Ω 0.00Ω~ 19.99Ω 10mΩ 10mA within ±0.2%rdg±2digit
Open-circuit voltage of measuring terminal: less than 5V
Measuring method: 4-terminal measurement
Sampling time:
[Free running mode] 12.5 times per second (Power 50Hz), 15 times per second (Power 60Hz)
[Remote start mode] 80msec. (Power 50Hz), 66.7msec. (Power 60Hz)
Comparator set range: +000~+1999 both for high and low limit
Indication of comparator's comparison result: LED indication LO/GO/HI and buzzer
Control signal:
Remote start input
Remote start input comparison output[LO/GO/HI]
Relay contact output:AC110V max.0.3A/DC20V max.1A
End of comparison output[EOC]:open collector: 50V/100mA(max.)
Power supply: AC100V~240V selectable, 50/60Hz, about 21VA
Outer dimension: about 260 (W) ×90 (H) ×250 (D) mm (excluding protruding parts such as rubber legs, etc.)