Measuring range and Accuracy (at 23*C±5*C)
Range / Measuring range / Measuring current / Accuracy Indication range / Indication range
1Ω / 0.01Ω~ 0.999Ω / 200mA / within ±α±2digit (SLOW) - within ±β±(3+n) digit (FAST) / ±9.99%
10Ω / 1Ω~ 9.99Ω / 100mA / within ±0.03%±1digit (SLOW) - within ±0.03%±(2+n) digit (FAST) / ±9.99%
100Ω / 10Ω~ 99.9Ω/ 10mA / within ±0.02%±1digit (SLOW) - within ±0.02%± (2+n) digit (FAST) / ±9.99%
1kΩ / 100Ω~ 999Ω/ 10mA / within ±0.02%±1digit (SLOW) - within ±0.02%± (2+n) digit (FAST) / ±9.99%
10kΩ / 1kΩ~ 9.99kΩ / 1mA / within ±0.02%±1digit(SLOW) - within ±0.02%± (2+n) digit (FAST) / ±9.99%
100kΩ / 10kΩ~ 99.9kΩ Ω / 100μA / within ±0.02%±1digit (SLOW) - within ±0.02%± (2+n) digit (FAST) / ±9.99%
1MΩ / 100kΩ~ 999kΩ / 10μA / within ±0.02%±1digit (SLOW) - within ±0.1%± (2+n) digit (FAST) / ±9.99%
10MΩ /1MΩ~9.99MΩ / 1μA / within ±0.04%±1digit (SLOW) - within ±0.4%± (2+n) digit (FAST) / ±9.99%
100MΩ / 10MΩ~99.9MΩ / 0.1μA / within ±0.2%±2digit ( SLOW) / ---------
/ - / mΩ CHECK(0~999mΩ) / 50mA / within ±0.15%±1digit(SLOW) - within ±0.2%±2digit (FAST) / 0~999mΩ
Measuring time
Power source frequency: 60Hz
Remote start:
SLOW: about 16.7msec.
FAST-1 (F1): about 2.1msec.
FAST-2 (F2): about 4.2msec.
Free running:
SLOW: 30 times per second
FAST: 30 times per second
Power source frequency: 50Hz
Remote start:
SLOW: about 20.0msec.
FAST-1 (F1): about 2.5msec.
FAST-2 (F2): about 5msec.
Free running:
SLOW: 25 times per second
FAST: 25 times per second
Open-circuit voltage of measuring terminal: less than 15V
Measuring method: 2-terminal measurement (more than 100kΩ) 4-terminal measurement (less than 99.9kΩ)
Setting range for measurement value:
±00.0%~±99.9% both for high and low limit In a case of mΩCHECK:000mΩ~999mΩ
Operation condition: [Temp.] 5℃~+40℃ [Humidity] less than 85%
Power supply: AC100V~240V switchable, 50/60Hz, about 50VA
Outer dimension: about 333 (W) ×85 (H) ×245 (D) mm (excluding protruding parts such as rubber legs, etc.)
Weight: about 3.5kg