Measuring range and Accuracy (at23*C±5*C) *D<0.5 parallel equivalent circuit
Range/ Measuring range/ Measuring voltage/ Accuracy
0/ 0.00~ 19.99pF/ about 1V (r.m.s.)/ within ±0.3%of rdg ±3digit
1/ 0.0~ 199.9pF/ about 1V (r.m.s.)/ within ±0.3%of rdg ±2digit
2/ 0.000~ 1.999nF/ about 1V (r.m.s.)/ within ±0.3%of rdg ±2digit
3/ 0.00~ 19.99nF/ about 1V (r.m.s.)/ within ±0.3%of rdg ±2digit
4/ 0.0~ 199.9nF/ about 1V (r.m.s.)/ within ±0.3%of rdg ±2digit
5/ 0.000~1.999μF/ about 1V (r.m.s.)/ within ±0.3%of rdg ±2digit
6/ 0.00~19.99μF/ about 1V (r.m.s.)/ within ±0.3%of rdg ±2digit
Measuring frequency: within 1kHz ±0.1%, SINE wave
C offset range: about 10pF
Sampling time:
[Remote start mode] 12.2msec.
[Free running mode] 45 times/sec.
Maximum displayl: 1999
External bias: max.DC25V
Comparator set range: 0000~1999 both HI and LO
Indication of comparator's comparison result: LED indication -NG, GO, +NG and buzzer
Control signal:
Remote start input:"L" [OV] →"H" [DC12V] start
Remote hold input: Open and "H" [DC12V]
:Free run/"L" [OV] :Hold
Comparison output [-NG/GO/+NG]:open collector:max.50V, 100mA
Completion of the measurement signal:EOC: open collector:max.50V, 100mA
BCD measurement data・Termination signal:Fan-out 2
Operation condition [Temp.]: +5*C~+40*C [Humidity] less than 85%
Power supply: AC100V~240V selectable, 50/60Hz, about 20VA
Outer dimension: about 260(W)×90(H)×250(D)mm (excluding protruding parts such as rubber legs, etc.)
Weight: about 2.6kg