- Temperature Range:
50 ~ 1000°C (122 ~ 1832°F) (MB 10)
75 ~ 1300°C (167 ~ 2372°F) (MB 13)
150 ~ 1800°C (302 ~ 3272°F) (MB 18)
- Spectral Ranges: 2 ~ 2.6 μm (main wavelength 2.3 μm)
- Resolution:
0.1°C or 0.2°F at interface
< 0.0015% of adjusted temperature range at analog output, 16 bit; 1°C or 1°F on display
- Emissivity ε: 0.050 ~ 1.000 in steps of 1/1000
- Transmittance τ: 0.050 ~ 1.000 in steps of 1/1000
- Response Time t90: 0.5 ms; (with dynamic adaption at low signal levels)
Adjustable to: 1 ms, 3 ms, 5 ms, 10 ms, 50 ms, 250 ms, 1 s, 3 s, 10 s
- Measurement Uncertainty (ε = 1, t90 = 1 s, Tamb. = 25°C):
< 1500 °C: 0.3% of reading in °C + 2°C
> 1500°C: 0.6% of reading in °C
- Repeatability (ε = 1, t90 = 1 s, Tamb. = 25°C):0.15% of reading in °C + 1°C
Optical Specifications:
- Sighting: Built-in laser aiming light (max. power level < 1 mW, λ = 630 ~ 680 nm, CDRH class II) or through-lens sighting
- Optics: Manually focusable from rear cover measuring distance a = 210 ~ 5000 mm
- Distance Ratio:
MB 10: approx. 50:1
MB 13: approx. 100:1
MB 18: approx. 350:1
Electrical:
- Power Supply: 24 VDC ±25%, ripple must be less than 50 mV
- Power Consumption: Max 3 W (incl. laser)
- Load (analog output): 0 ~ 500 Ω
Environmental
- Protection Class: IP 65 IEC 60529 (value in mated condition)
- Ambient Temperature: 0 ~ 70°C (32 ~ 158°F)
- Storage Temperature: -20 ~ 80°C (-4 ~ 176°F)
- Weight: 0.6 kg
- Housing: Stainless steel
- CE Label: According to EU directives about electromagnetical immunity
Interface:
- Connection: 12-pin connector
- Display (in rear cover): LED, 4 digit matrix, 5 mm high temperature signal or measuring distance
Communication
- Analog Output: 0Adjustable 0 ~ 20 mA or 4 ~ 20 mA, linear (via digital interface)
- Digital Interface: RS485 addressable (half-duplex)
Baud rate: 1200 Bd to 115.2 kBd (on request RS232, not addressable)
- Scope of Delivery: Pyrometer with PC software InfraWin for adjustment and evaluation, Works Certificate, and Manual