MAXIMUM INPUT RATINGS
Current : 24A
Voltage : 120V
Power : 120W
Min operating voltage : ≤ 700mV @ 24A
Min effective resistance : 30mΩ (typical)
Off state leakage : <60µA (typical)
Reverse Polarity : 24A (diode will conduct)
Isolation Voltage : ±300Vdc to chassis ground
OPERATING MODES
CC MODE
Ranges : 0 to 2A (limited to 40W). 0 to 24A
Accuracy : ±(0.1% + 0.03% Range)
Resolution : 2A range: 100uA. 24A range: 1mA
Regulation : Same as accuracy specification (auto-compensating)
Minimum Transition time at full current : 2A range: 4µs. 24A range: 12µs
CP MODE
Range : 0 to 40W (limited to 2A). 0 to 120W
Accuracy : ±(0.2% + 2mW + 0.03% I Range)
Resolution : 1mW
Regulation : Same as accuracy specification (auto-compensating)
Minimum Transition time at full current : 40W range: 4µs. 120W range: 12µs
CR MODE
Range : 100mΩ to 10kΩ
Accuracy : ±(0.2% + 2mΩ + 0.03% I Range)
Resolution : 4 digits (minimum 1mΩ)
Regulation : Same as accuracy specification (auto-compensating)
Minimum Transition time at full current : 2A range: 4µs. 24A range: 12µs
CV MODE
Range : 0 to 120V
Accuracy : ±(0.05% + 0.01% FS)
Resolution : 1mV
Regulation : Same as accuracy specification (auto-compensating)
INTERFACES
USB : Full digital remote control facilities are available through the USB interface. Standard USB 2.0 hardware connection. Implemented as virtual-COM port. SCPI compatible
Remote Command Processing Time : Typically < 100ms
GENERAL
Terminals : 4mm Banana sockets on 19mm spacing
Display : 2.8” IPS TFT (320x240) Backlit, 4W resistive touch
Data Entry : Menu navigation via touchscreen or rotary knob, value entry direct by numeric keys or by rotary control
Stored Settings : Up to 6 complete instrument set-ups may be stored in non-volatile memory
Size and Weight : 213.3 x 227.6 x 98.2 mm (WxDxH), 1.7kg
Power : 100-240VAC ±10% 50/60Hz; 30VA max. Installation Category II
Operating Range : +5°C to 40°C, 20-80% RH
Storage Range : -20°C to +60°C
Cooling : Variable Speed fan. Air exit at rear.
Environmental : Indoor use at altitudes up to 2000m, Pollution Degree 2
Safety & EMC : Complies with EN61010-1 & EN61326-1. All inputs 300Vpk max to ground.