Withstanding voltage
• Output voltage:
Range: (300 ~ 3000) VAC; Step: 10V
Accuracy: ± (2.5%x setting +10V); Distortion: <3%
• Output frequency:50Hz/60Hz; Accuracy: ±0.1Hz
• Actual capacity: ≥ 90%
• Breakdown current:
Range: (0.10~20.00) mA
Accuracy: ± (2%x reading +5 counts)
• Duration: Range: (1~99) s; Accuracy: ± (1 %x setting
+0.1s)
• Arc: Level- 0 to 9, 0: Arc Off
Insulation Resistance
• Output voltage:
Range: (200 ~ 1000) V DC; Step: 5V
Accuracy: ± (3%x setting +5V)
• Ripple coefficient: <5%
• Resistance measurement:
Range: (1.0 ~ 99.9) M Ω, (100 ~500) M Ω
Accuracy: ± (3%x reading +5 counts), ≤200M Ω; ± (5%x reading
+8 counts), >200M Ω
• Duration: Range: (1~99) s; Accuracy: ± (1 %x setting
+0.1s)
Impulse
• Pulse voltage: Range: (500~3000) V; Step: 10V; Accuracy: ±
(3%x setting +10V)
• Storage capacitance: Standard capacitance: 0.022 μF
• Wavefront time: ≤0.5 μs
• Waveform acquisition:
Frequency: 5kHz~40MHz, 13-level adjustable.
Resolution: 8-bit AD
Storage: 1k bytes.
Multiple tests: (1 ~ 10) times continuous
• Waveform parameter:
Area, difference area, range: (0 ~ 99)%
Resolution: 0.1%; Accuracy: ±3.0%
Resistance
• Resistance measurement:
Range: 0.1~0.2/ 2/20/200/2k/20k, unit: Ω
Accuracy: ±(0.3%x reading +3 counts) (temperature compensation
off)
• Temperature measurement: Range: (0.0~60.0)°C; Accuracy:
±0.5°C
Inductance
Measurement range: 0mH~200mH
Resolution: 0.001mH
Measurement frequency: 100Hz/1000Hz
Measurement accuracy: ±(0.5% x reading + 4 counts)
Back EMF
Measurement range: 1~100VAC
Measurement accuracy: ±(0.5% x reading + 4 counts)
Hall waveform
(speed)/FG waveform (speed)
Voltage range: (0.50~10.0)V
Voltage accuracy: ±(0.5% x reading + 2 counts)
Frequency range: (10.0~1000.0Hz)
Frequency accuracy: ±(0.1% x reading + 1 count)
Speed range: (frequencyx60) RPM
Duty cycle: Range: (20~80)% Accuracy: ±1%
(absulote error)