Grounding resistance test
-Rated output
+ Maximum output current: 32A
+ Maximum test resistance: 600mΩ
+ Open circuit voltage less than 12V
-Output current setting: 2.0-32.0A, ±(2%xsetting value +5 counts)
-Output voltage: 3.0-10.0V, ±(2%xsetting value +5 counts)
-Output frequency: Sine wave, 50Hz or 60Hz ±1%xsetting value
-Range of resistance:
2.0-10.6A: 0.1-600mΩ;
10.7-32.0A: 0.1—RmΩ; R=(6400/setting current value)
-Time range: 0.1-999.9s;±(0.2%xsetting value+1 count)
-Communication port: Standard, network port, RS232, RS485
-Power supply: 220V±10%, 50Hz±5%
-Overall weight: About 25kg
-Dimension: 426(W)x133(H)x550(D)