Maximum temperature: 95 °C (203 °F) at 20 °C (68 °F) ambient temperature, no cooling
Viscosity range with DMA M
Operation with internal air pump: Up to 10,000 mPa·s at 95 °C heating and measuring temperature
Operation with 2 bar external compressed air: Up to 36,000 mPa·s at 95 °C heating and measuring temperature
Viscosity range with SVM x001: 0.2 mPa·s to 5,000 mPa·s at 95 °C heating and measuring temperature
Filling mode: Pressure mode:
with integrated piston pump
with air pump of the master instrument
with compressed air
Minimum sample volume per measurement: Approx. 4.5 mL with DMA M
Approx. 6 mL with SVM X001
Number of sample vials
Magazine, heated positions only: 36 positions for 12 mL vials (only glass vials allowed!)
Xsample heated / unheated positions: 32 positions heated / 24 positions unheated for 12 mL vials (only glass vials allowed!)
Automatic filling and drying: Internal air pump of the master instrument or compressed air selectable
Compressed air supply (optional): Max. 2.0 bar (29 psi) relative
Quality of air necessary: Class 3 according to ISO 8573-1:
max. particle size: 5 µm
max. pressure dew point: -20 °C (-4 °F)
max. oil content: 1 mg/m3
Environmental conditions (EN 61010): Indoor use only
Ambient temperature: 15 °C to 35 °C (59 °F to 95 °F)
Air humidity: 10 % to 90 % relative humidity, non-condensing
Pollution degree: 2
Overvoltage category: II
Altitude: Max. 3,000 m (9,840 ft)
Max. temperature of heated parts in case of malfunction: 160 °C (320 °F)
Power supply: Power is supplied by the master instrument and the Xsample heating controller