- Frequency Characteristics:
Sine: 1 µHz to 120 MHz
Square: 1 µHz to 25 MHz
Triangle, Ramp: 1 µHz to 1 MHz
Pulse: 1 µHz to 25 MHz
Gaussian Noise (-3 dB): > 120 MHz
Arbitrary: 1 µHz to 20 MHz
Accuracy: ± 1 ppm (1 year)
Resolution: 1 µHz
- Arbitrary Characteristics:
Built-in Waveforms: 196
Waveform Length: 8 points to 8 M points
Vertical Resolution: 16 bits
Sampling Rate: 300 MSa/s (DDS mode); 75 MSa/s (true arbitrary mode)
Minimum Rise/Fall; Time (typical): 4.5 ns (DDS mode); 8.5 ns (true arbitrary mode)
Jitter (rms): < 150 ps (1 Vpp, into 50 Ω load, true arbitrary mode)
Non-volatile Memory Storage: 80 MB fle system
- Output Characteristics:
Amplitude Range(1) (into open circuit):
2 mVpp to 20 Vpp ( ≤ 20 MHz)
2 mVpp to 10 Vpp ( > 20 MHz)
Amplitude Resolution: Up to 4 digits
Amplitude Accuracy (10 kHz, 0 V offset): ± (1% + 1 mVpp)
Amplitude Flatness (reference to 10 kHz Sine, 2.5 Vpp):
± 0.3 dB (50 Ω load, DC to 100 MHz)
± 0.4 dB (50 Ω load, 100 MHz to 120 MHz)
Cross Talk: < -60 dBc (between channels)
Offset Range (DC):
± 5 V (into 50 Ω load)
± 10 V (into open circuit)
Offset Resolution (DC): Up to 4 digits
Offset Accuracy (DC): ± (1% + 2 mV), into open circuit
Output Impedance (typical): 50 Ω
Output Protection: Overvoltage (see user manual for details)
- Waveform Characteristics:
Harmonic Distortion (sine, 0 dBm input, typical)
DC to 10 MHz, < -65 dBc
10 MHz to 20 MHz, < -60 dBc
20 MHz to 40 MHz, < -55 dBc
40 MHz to 60 MHz, < -50 dBc
60 MHz to 80 MHz, < -45 dBc
80 MHz to 100 MHz, < -40 dBc
100 MHz to 120 MHz, < -38 dBc
Total Harmonic Distortion (sine): < 0.075% (10 Hz to 20 kHz at 0 dBm)
Spurious (non-harmonic):
≤ 50 MHz, -70 dBc max.
> 50 MHz, -65 dBc max.
Rise/Fall Time (square): < 9 ns (10% to 90% at 1 Vpp, into 50 Ω load)
Variable Duty Cycle (square): 0.001% to 99.999% (depending on frequency setting)
Jitter (rms) Cycle to Cycle (square): 150 ps (1 Vpp, into 50 Ω load, typical)
Ramp Symmetry: 0% to 100%
Ramp Linearity: < 1% of peak output (triangle, ramp at 1 kHz, 1 Vpp, 100% symmetry)
- Environmental and Safety:
Temperature Operating: 32 °F to 104 °F (0 °C to 40 °C)
Storage: -4 °F to 140 °F (-20 °C to 60 °C)
Humidity: < 86 °F (30 °C), ≤ 90 % RH; 104 °F (40 °C), ≤ 50 % RH
Altitude: Operating: below 10,000 ft (3,048 m); Storage: below 49, 212 ft (15,000 m)
Electromagnetic Compatibility: EMC Directive 2014/30/EU, EN61326-1:2013
Safety: Low voltage directive (LVD) 2014/35/EU, EN61010-1:2010
- General:
Display: 4.3” TFT color (24-bit) LCD touch screen
I/O Interfaces: USBTMC device, LAN, USB host port
Storage Memory: 10 instrument settings
AC Input: 100 to 240 VAC ± 10 %, 50/60 Hz; 100 to 120 VAC ± 10 %, 400 Hz
Power Consumption: 50 W maximum
Dimensions (W x H x D): 10.25” x 4.22” x 11.61” (260.3 x 107.2 x 295 mm)
Weight: 7.6 lbs (3.43 kg)
- Standard Accessories:
AC power cord, user manual (downloadable),
USB type A-to-B cable, BNC coaxial cable,
certifcate of calibration
- Optional Accessories: USB-to-GPIB adapter (model AK40G)