EPC(electronic pneumatic control)
7 inch touch screen display
Sample Introduction System: sample injector and evaporation chamber
Column Oven:
Temperature range: Ambient temperature +7℃ ~ 420℃ (in 1℃ increment)
Temperature accuracy: ± 0.02℃
Cooling time: 350℃~100℃≤3min, 400℃~50℃ in 8-10 min at 25℃ ambient
Programmed temperature setting: 0.1℃~80℃/min (in 1℃ increment)
Program ramps: 20 ramps in total (99 ramps available with control workstation)
Size (L*W*H): 284*280*241mm(inside)
Flame Ionization Detector (FID):
Detection limit: ≤8×10-12g/s (n-hexadecane)
Drift: ≤1.5×10-13A/h
Noise: ≤3×10-14A
Linearity range: ≥106
Flame Photometric Detector (FPD):
Detection limit: ≤8×10‐13g/s(P) ; ≤8×10‐11g/s(S)
Drift: ≤2×10‐11A/30min
Baseline noise: ≤5×10‐12A
Nitrogen-Phosphorus Detector (NPD):
Detection limit: ≤5×10-12g/s(N) (Azobenzene); ≤5×10-13g/s(P) (Malathion)