Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Measuring Range
Rangeability: 10:1 (most sizes)
Metering Tube
Standard: Tefzel® Lined 316/316L (dual certified stainless steel)
Flanges and End Fittings
Standard: Tefzel® Lined 316/316L (dual certified stainless steel)
Accuracy: 2%, VDI/VDE class 2.5
Repeatability: 0.25% Full Scale
Scale type/material: Dark increments with white background/Aluminum
Installation orientation and location: Vertical (within 5% of true vertical), bottom inlet, top outlet. Do not locate in proximity of other magnetic interfering components
Connections
Flanged: Slip on flanges
Equivalent to ANSI B16.5: ANSI 1/2” to 2” 150# RF to 300# RF
DIN 2527/ EN 1092-1: DIN PN 40
Flange finish: 3.2 - 6.3 Ra
Floats
Standard: Hastelloy C - 276 (sizes 7.8) PVDF (sizes 10 - 13)
Protection Category
Indicator only: IP66/67 NEMA 4X
Transmitter ALU: IP66/67 NEMA 4X**
Transmitter SS: IP66/67 NEMA 4X
Indicator Housing & Cover material
Indicator only ALU: Die cast Aluminum (Alloy 380), epoxy paint, glass window
Transm/Alarm/HiTemp ALU: Die cast Aluminum (Alloy 380), epoxy paint, glass window
Indicator only SS: Cast 316L stainless steel, glass window
Transm/Alarm/HiTemp SS: Cast 316L stainless steel, 316 stainless steel hardware, glass window
Pressure/Temperature
Maximum Fluid Temperature: 150°C/270°F
For more details, see datasheet.