HUMAN-MACHINE INTERFACE
Type of display: 3.5” colour TFT – Resolution 320x240 –LED backlighting
Display mode: 2,500 real acquisition points on screen
On-screen display of curves: 2 curves + 2 references + memory trace or mathematical calculation
Commands: Direct adjustments on front panel & on-screen menus via browser (main & secondary without «hidden menus»)
Interactive help function: 14 languages: French, English, German, Spanish, Italian, Swedish, Romanian, Russian, Finnish, etc.
OSCILLOSCOPE MODE
VERTICAL DEFLECTION
Bandwidth: 20 MHz
Bandwidth limieter: 1,5 MHz, 5 kHz
Number of channels: 2 totally-isolated channels
Input impedance: 1 MΩ ±0.5%, approx. 17 pF
Maximum input voltage: 600 V CAT III – Derating -20 dB per decade from 100 kHz
Vertical sensitivity: 5 mV to 200 V/div
HORIZONTAL DEFLECTION
Sweep speed: 25 ns/div to 200 s/div –Roll mode from 100 ms to 200 s/div
Horizontal zoom: Zoom coefficient: x1, x2, x5
TRIGGERING
Mode: Automatic, triggered, one-shot & Triggered Roll
Type: Edge, pulse width (20 ns – 20 s)
Coupling: AC or DC (depending on coupling of triggering channel)), HF, LF or noise rejections
Sensitivity: ≤1,2 divisions p-p up to 40 MHz
DIGITAL MEMORY
Maximum sampling rate: 2 GS/s in ETS mode – 50 MS/s in one-shot mode on each channel
Vertical resolution: 9 bits
Memory depth: 2,500 points per channel
User memory: 2 MB for file storage: trace (.trc), text (.txt), configuration (.cfg), image files (.bmp)
GLITCH mode: Duration ≥20 ns – 1,250 Min/Max pairs
Display modes: Envelope, Averaging (Factors 2 to 64) and XY (vector)
OTHER FUNCTIONS
MATH functions Channel inversion, addition, subtraction, multiplication and division (adjustable scaling)
Cursor measurements 2 cursors: V, T, dV, dt simultaneously – 4-digit display resolution
Automatic measurements 18 time-based or level measurements and Phase measurement
MULTIMETER MODE
General specifications: 2 channels, 8,000-count display + min/max bargraph – Graphical recording of 2,700 measurements (5 min to 1 month)
Operating modes: Absolute or relative display (absolute, deviation, rel, rel%) – Monitoring (instantaneous, Min, Max, Avg)
AC, DC and AC+DC voltages: Ranges From 600 mV to 600 VRMS, 800 mV to 800 VDC –VDC Accuracy 1% Reading +20D –50 kHz bandwidth
Resistance: Range from 80 Ω to 32 MΩ - accuracy 2%R+10D –10 ms quick continuity test
Capacitance: Ranges from 5 nF to 5 mF – basic accuracy 2%R+10D
Other measurements: Frequency, rotation speed, 3.3 V diode test, temperature measurement (by K Thermocouple and infrared sensor)
POWER
Measurements: Single-phase and balanced three-phase active power values (with or without neutral), simultaneous display of current – PF
HARMONIC ANALYSER MODE
Multi-channel analysis: 2 channels, 31 orders, frequency of fundamental from 40 to 450 Hz
Simultaneous measurements: Total VRMS, THD and selected order (%fundamental, phase, frequency, VRMS)
GENERAL SPECIFICATIONS
Screenshots: Up to 100 files in standard «.bmp» format, viewable on the instrument
PC communication: Isolated optical USB interface – SX-Metro PC application available as an option + SX-DMM for multimeter mode
Power supply: 6 x LR6 batteries or 6 x AA NiMH rechargeable batteries – Battery life of up to 8.5 hrs
Safety / EMC: Safety as per IEC61010-1 Ed3 – 600 V CAT III – EMC as per EN61000-3, 2001 & EN61326-1, 2006