How common is the selection of RBW on a spectrum analyzer?

11/10/2026 15:13:05

You are measuring an amplifier, and only one peak appears on the screen—even though there are clearly two carriers. Or, you know for a fact that harmonics are present, but the noise floor sits above them. Or, after making adjustments, the scanner slows down to the point where you find yourself waiting for each individual sweep. These three scenarios often share a common cause: the RBW (Resolution Bandwidth)

You measure an amplifier and the screen shows a single peak, even though you are sure there are two carriers. Or you know a harmonic is there, but the noise floor sits above it. Or you finish adjusting and the analyzer sweeps so slowly that you wait on every pass. These three situations usually share one cause: RBW.

Choosing RBW on a spectrum analyzer

What RBW is, in brief

A spectrum analyzer does not look at the whole frequency range at once. It sweeps a narrow filter (the intermediate-frequency, or IF, filter) across the range and plots the power at each point. The width of that filter is the RBW (Resolution Bandwidth).

Think of RBW as the width of the "window" the analyzer listens through at each step. A narrow window lets you hear each frequency separately, but you have to move slowly. A wide window moves fast, but everything inside it is merged into one.

Adjusting RBW changes three things at once

ParameterSmall (narrow) RBWLarge (wide) RBW
Separating close signalsGoodPoor, peaks merge
Displayed average noise level (DANL)LowerHigher
Sweep speedSlowFast
The 10 dB rule. Each time you reduce RBW by a factor of 10, the displayed noise floor drops by about 10 dB. The real signal does not change, so it rises another 10 dB above the noise.
RBW100 kHz10 kHz1 kHz100 Hz10 Hz
Noise floor (relative to reference)reference-10 dB-20 dB-30 dB-40 dB
The price you pay is sweep time. In normal swept mode, sweep time is proportional to the Span divided by the square of the RBW. Reducing RBW by a factor of 10 can make the sweep up to 100 times longer. Some analyzers offer an FFT mode for narrow RBW, which is less slow in practice, but do not count on it when the Span is very wide.

Five measurement situations and how to set RBW

01  Separating two closely spaced carriers

Two carriers at 900.00 MHz and 900.01 MHz, 10 kHz apart. This is what you will see on the screen:

RBW 30 kHz  ▁▂▄▆██▆▄▂▁  → only 1 peak
RBW 3 kHz   ▁▃▇▃▁▁▃▇▃▁  → splits into 2 peaks

Setting: RBW less than or equal to the spacing between the two signals. To make the two peaks clearly separate, choose about one third of that spacing, which is 3 kHz in this example.

Note: if the two signals differ a lot in amplitude, the smaller one can hide under the "skirt" of the larger one, because the filter is not infinitely steep. In that case you need an even smaller RBW.

02  Finding very small harmonics and spurious signals

Setting: first narrow the Span around the suspected frequency, then reduce RBW step by step (for example 10 kHz, 1 kHz, 100 Hz, 10 Hz). Each step lowers the noise floor by 10 dB, so any spurious signal buried in the noise gradually emerges.

Note: if you reduce RBW but leave the Span wide, the sweep time becomes very long. Do it in order: narrow the Span first, then reduce RBW.

03  Seeing sidebands when measuring AM signals

A 100 MHz carrier modulated by a 1 kHz signal has two sidebands at ±1 kHz from the carrier.

Setting: RBW below the modulating frequency, about one third of it or lower. For 1 kHz, choose 300 Hz or less, otherwise the sidebands merge into the carrier.

04  Measuring channel power of a wideband signal (LTE, 5G NR, Wi-Fi)

A digitally modulated signal has a spectrum like a flat block of noise, several MHz wide. You do not need an extremely narrow RBW here, but you should not let RBW get too large relative to the channel either.

Setting: RBW of about 1 to 3% of the channel bandwidth. For example, for a 20 MHz channel choose about 200 to 600 kHz. The easiest way is to use the Channel Power measurement function, which sets a suitable RBW and detector for you.

Note: an RBW that is too large relative to the channel rounds off the channel edges and affects adjacent channel power (ACPR) results.

05  Hunting for unknown signals across a wide frequency range

Setting: wide Span and large RBW (about 100 kHz to 1 MHz) so the analyzer sweeps quickly. When you see a suspicious peak, place a marker on it, narrow the Span around it, and only then reduce RBW to look closely. This is the most time-efficient approach.

Read more:
RF measurement distorted or worried about burning the probe: when to switch to a signal analyzer

Quick reference table

Measurement goalSuggested RBWCombine with
Hunting for signals across a wide range100 kHz to 1 MHzWide Span, fast sweep
Separating two carriersAbout 1/3 of the spacing between themSpan just wide enough for both
Finding small harmonics and spurious signalsStep down from 10 kHz to 10 HzNarrow the Span first
Modulation sidebands1/3 of the modulating frequency or lowerSpan of a few times the modulating frequency
Wideband channel powerAbout 1 to 3% of the channel bandwidthUse the Channel Power function

VBW: it smooths the trace, it does not lower the noise floor

Next to RBW there is always VBW (Video Bandwidth), a post-detector filter used to smooth the displayed trace. On many Keysight analyzers, VBW defaults to the same value as RBW (a 1:1 ratio).

When the noise floor looks jagged like a saw blade, you can lower VBW (VBW/RBW = 0.1 or 0.01) to make the trace smoother and the peaks easier to read. But keep these points clear:

VBW only smooths. It does not lower the average noise floor and does not separate any additional signals.
Lowering VBW too far below RBW also slows the sweep. Lower it only as much as you need to read the data.
To get a truly lower noise floor: reduce RBW. To get a smooth trace without a long wait: use averaging (averaging over several sweeps).

Common mistakes when adjusting RBW

1Setting a very low RBW but leaving a wide Span. The analyzer seems to freeze. Narrow the Span first.
2Ignoring the "Meas Uncal" warning. This message means the analyzer is sweeping too fast for the RBW, so the measured amplitude can no longer be trusted. Leave Sweep Time on Auto.
3Signal skirts spreading wide at a very narrow RBW. This is usually phase noise from the source or from the analyzer itself, not necessarily a spurious signal.
4Peaks jumping around at a very narrow RBW. This is usually caused by the source drifting in frequency, and the analyzer is not necessarily at fault.

Choosing RBW in four steps

Step 1. Set the Span to the frequency region you need to see.

Step 2. Turn RBW to Auto. The analyzer typically sets RBW to about 1/100 of the Span, enough to balance resolution and sweep speed.

Step 3. Decide your goal (separating signals, lowering the noise floor, or measuring channel power), switch to Manual, and set RBW from the quick reference table.

Step 4. Check the result: no Meas Uncal warning and stable peaks. If the sweep is too slow, narrow the Span first, then relax RBW.

Need a spectrum analyzer that fits your measurement?
EMIN supplies Keysight signal analyzers (N9000B, N9010B, N9021B) and FieldFox handheld analyzers for field measurements. Send EMIN's engineers your frequency range, signal type and the power level you need to measure, and we will advise on the right configuration and settings.

Related News

Vietnam’s Semiconductor Industry Outlook to 2030: Opportunities and Challenges
11/10/2026 11:22:48

Những khoản đầu tư mới, cùng định hướng phát triển công nghiệp bán dẫn đến năm 2030, đang tạo nền tảng cho thị trường trong nước. Tuy nhiên, để nâng cao vị thế trong chuỗi giá trị toàn cầu, Việt Nam cần giải quyết bài toán về công nghệ lõi, nhân lực kỹ thuật và năng lực đo kiểm.

Choosing Fans, Heatsinks, Heat Spreaders, or Thermal Paste for High-Performance Electronic Devices
11/09/2026 13:50:49

SiC and GaN semiconductor components feature high power density and compact dimensions, causing heat to be concentrated within a very small chip area. The question of which thermal management solution is most widely used is often framed as a choice between four options: fans, heatsinks, thermal pads, or thermal grease. This framing requires some refinement, as these four components occupy different positions along the same heat-transfer path; in fact, most designs employ a combination of several of these elements. This article explains the role of each component, how to select thermal interface materials, and the impact of temperature on measurements.

EMIN GROUP AT VIMF BAC NINH 2026
11/07/2026 10:52:42

EMIN GROUP participated in the Vietnam Industrial & Manufacturing Fair 2026 (VIMF Bac Ninh 2026), held from November 4 to 6, 2025, at Kinh Bac Cultural Center in Bac Ninh City.

How Do Bidirectional DC Power Supplies Improve EV Testing?
11/07/2026 09:57:58

Bidirectional DC power supplies combine source and sink operation in one platform. This supports realistic EV battery cycling, regenerative energy handling, programmable test sequences, and flexible battery or charger emulation.

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts