CHROMA 19032 Electrical Safety Analyzer (6kVDC, 5kVAC)
Mode : ACWV/DCWV/IR/GB/LC
Withstanding Voltage Test
Output Voltage : DC: 0.05-6kV; AC: 0.05-5kV
Load Regulation : ± (1% of reading +5V)
Voltage Regulation : 2V
Voltage Accuracy : ± (1% of reading + 0.1% of full scale)
Cutoff Current : DC: 12mA; AC: 40mA
Current Resolution : 0.1 μA DC; 1 μA AC
Current Accuracy : ± (1% of reading + 0.2% of full scale)
Output Frequency : 50Hz/60Hz
Test Time : 0.3-999 sec., continue
Ramp Time : 0.1-999 sec., Off
Fall Time : 0.1-999 sec., Off
Waveform : Sine wave
Insulation Resistance Test
Output Voltage : DC: 0.05-1kV
Voltage Resolution : 2V
Voltage Accuracy : ± (1% of reading + 0.5% of full scale)
IR Range : 0.1MΩ-50GΩ
Resistance Resolution : 0.1MΩ
Resistance Accuracy : 5% typical
Ground Bond Test
Output Current : AC: 1-30A
Current Accuracy : ± (1% of setting + 1% of full scale)
GR Range : 10mΩ-510mΩ
Resistance Resolution : 0.1mΩ
Resistance Accuracy : ± (1% of reading + 0.1% of full scale)
Test Method : 4 wires
Flashover Detection
Setting Mode : Programmable setting
Detection Current : AC: 20mA; DC: 10mA
Secure Protection Function
Ground Fault Interruption : -
Floating Output to Ground : -
Panel Operation
Lock : Present password
Interlock : YES
GO/NG Judgment Window
Indication/Alarm :
GO: Short sound, Green LED
NG: Long sound, Red LED
Data Hold
Memory Storage : 50 setups with up to 100 groups recall
Interface
Interface : 9 pin D-sub I/O control / RS-232 / GPIB (optional)
General
Operation Environment : Temperature: 0°C-40°C, Humidity: 20%-80% RH
Power Consumption :
No load: < 100 W
Rated load: 800 W
Power Requirements : 90-132Vac or 180-264Vac, 47-63Hz
Dimensions (H × W × D) : 133 × 430 × 470 mm / 5.24 × 16.93 × 18.66 inch
Weight : 25.5 kg / 56.17 lbs
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts