Temperature range: 50℃ ~ 600℃
Calibration plane diameter: 130mm
Uniformity:
0.3℃ @RT ≤ t ≤ 100℃
0.7℃ @100℃ <t ≤ 300℃
1.2℃ @ 300℃< t ≤ 400℃
1.8℃ @ 400℃ <t ≤ 500℃
Stability (10min):
±0.2℃ / 10min @ RT ≤ t ≤100℃
±0.3℃ / 10min@100℃<t ≤ 300℃
±0.4℃ / 10min@300℃<t ≤ 400℃
±0.6℃ / 10min@400℃<t≤ 500℃
Resolution: 0.01℃
Temp. rising time: Room temperature to 400℃, 30min
Temp. falling time: 400℃ to room temperature (with rapid cooling device only 45min)
Stabilization time: 8min
Communication Interface: RS-232
Sensor type: RTD / Pt100
Heating furnace size: 325(L)×323(W)×214(H) mm
Controller size: 220(L)×175(W)×144(H) mm
Total Weight: 12kg
Storage condition: − 20℃ ~ 60℃ (- 4℉~ 140℉), 5%RH ~ 80%RH ( No condensation )