Measurements:
O2%:
Range: 0-25 %
Accuracy*: ± 0.2% Vol
Resolution: 0.01% Vol
Type: EC
Method: Sampled or on-demand
CH4%:
Range: 0-100%
Accuracy*: ± 0.3% Vol or 3 % m.v
Resolution: 0.01% Vol
Type: NDIR
Method: Sampled or on-demand
CO2%:
Range: 0-100%
Accuracy*: ± 0.3% Vol or 3 % m.v
Resolution: 0.01% Vol
Type: NDIR
Method: Sampled or on-demand
H2S:
Range: 0-5000 ppm
Accuracy*: ± 5 ppm
Resolution: 1 ppm
Type: EC
Method: Sampled or on-demand
Barometric Pressure: 850-1150 mbar
Calculated Values: CV gross/net 0-50 MJ3, MJ/Kg
Display: 3.5″ TFT color display, backlight
Keyboard: Tactile with password protected menu
Output communications/Industrial Interfaces:
Analouge: up to 12 x 4-20 mA outputs/input individually configurable
Alarms: 1 x system alarm and up to 6 alarm relay output , free contact 24 VDC/5A
Digital: Modbus RTU as standard, optional Ethernet, profinet, profibus DP and wed enable remote access
Data logging: Internal memory and SD card
Event Log: Record of Alarm/event to SD card
Sample:
Gas Inlet: Stainless steel gas gas fittings with 1/8″ ID Thread
Water Trap: Integrated water trap with auto-drain as standard
Filter: Teflon Particulate filter, internal housing in Viton
Flow: 40-60 l/hr
Pressure: Sample pressure from -100 to 300 mbar
Safety: LEL% internal gas sniffer
Sample gas shut-off insolent valve
Stainless Steel flow orifice restriction
Physical:
Dimension: 400x500x300 (H x W x D) wall mounting
Weight/Protection: 14 KG, IP 54
Operating Temperature: +5 to 45 °C ( -10 to 45 °C with optional heater)
Heater: Cabinet Heater 200 W
Mains/Power: Universal 90240 VAC/ 47..63 Hz, 42W – 242 W with cabinet heater
Notes:
*After Calibration- Does not include calibration gas accuracy
** Due to continuous product improvements, specifications might change with-out prior notice