Main Chamber Nominal Size (Width): 2 meters
Working Chamber Dimensions - Min (W x D x H): 2 m x 0.55 m x 0.77 m
Working Chamber Dimensions - Max (W x D x H): 2 m x 0.63 m x 0.84 m
External Dimension (with one Pass Chamber) (W x D x H):
With Adjustable Base Stand (Min): 2.62 m x 0.92 m x 2.55 m
With Adjustable Base Stand (Max): 2.62 m x 0.92 m x 2.85 m
External Dimension of Pass-through Chamber (W x D x H): 0.62 m x 0.66 m x 0.58 m
Glove Port Height (Min): 1200 mm
Glove Port Height (Max): 1500 mm
Chamber Environment: ISO Class 5 all chambers (Grade A)
Airflow:
Class III: Turbulent
Class I: Unidirectional (Positive)
Pre-filter:
Process Chamber: G4, polyester media
Pass-through Chamber: G4, polyester media
Filter Type - Chamber Inlet: HEPA (H14) Filter with Integral Mesh Guard and Gasket Seal
Filter Efficiency - Chamber Inlet: 99.995% MPPS
Filter Type - Chamber Exhaust - 1st Stage: HEPA (H14) Filter with Integral Mesh Guard and Gasket Seal
Filter Type - Chamber Exhaust - 2nd Stage: HEPA (H14) Filter with Integral Mesh Guard and Gasket Seal or Carbon Filter
Filter Efficiency - Chamber Exhaust - 1st Stage: 99.995% MPPS
Filter Efficiency - Chamber Exhaust - 2nd Stage: 99.995% MPPS
Lighting Level: ≥ 500 Lux (6000 K)
Sound Level: ≤ 80 dBA
Isolator Construction:
Chamber: SS 316L
Service Housing: SS 304
Support Frame: SS 304
Isolator Finish:
Chamber Internal: ≤ 0.4 Ra
Chamber External: ≤ 0.6 Ra
Service Housing External: ≤ 0.6 Ra
Support Frame: ≤ 1.0 Ra
Electrical Requirements: 220-240 VAC, 50/60 Hz, 1Ø
Compressed Air Requirement (by Client) If no on-board compressor: 2 Bar-g Pressure at 5 L/sec ✔
Exhaust Duct Requirements (by Client) unless Integral Catalytic Converter is Included: 10” Duct from Isolator to Outside
Net Weight: TBD