Fluke (Biomedical) QA-ES MK III-07 Electrosurgical Analyzer (SCHUKO, non-wireless)
Physical
Housing: Metal case
Size (H x W x L): 14.5 cm x 35 cm x 47 cm (5.75 in x 13.75 in x 18.5 in)
Weight: 7.5 kg (16.5 lbs)
Power
Power requirements: 100 V ac, 115 V ac, 230 V ac, 50 Hz / 60 Hz, universal input
100 V/115 V: 20 VA
230 V: 30 VA
User Interface
LCD: Monochrome 240 pixels x 64 pixels, 8 lines x 40 characters, white LED backlight
Keys: 6 (1 fixed, 5 soft-defined) and rotary selector knob
Environmental Specifications
Operating temperature: 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature: -20 °C to 60 °C (-4 °F to 140 °F)
Humidity: 10 % to 90 % non-condensing
Altitude: 2000 m maximum
IP rating: IEC60529:IP20
Measurements and Test Specifications
Measures: Cut and coag waveforms, monopolar and bipolar outputs
Power and current measurements: True-rms
Bandwidth: 30 Hz to 3 MHz at -3 dB including loads
Delay time for single measurements: 0.2 seconds to 4.0 seconds from Foot Switch activation to start of measurement
Duty Cycle
Variable load: 10 seconds on, 30 seconds off, at 100 W, all loads
Fixed 200 Ω load: 10 seconds on, 30 seconds off, at 400 W
Generator Output Measurements
Load resistance
Variable: 0 Ω, 10 Ω, 20 Ω, 25 Ω to 2500 Ω (by 25 Ω), 2500 Ω to 5200 Ω (by 100 Ω)
DC accuracy: ± 2.5 %
Power
Ranges: 0.0 W to 99.9 W; 100 W to 500 W
Accuracy:
< 50 W: +/- (6.25 % + 1 W)
≥ 50 W: +/- 6.25 %
Maximum:
At 25 % duty cycle (10 seconds on, 30 seconds off):
10 Ω: 300 W
20 Ω to 2900 Ω: 400 W
3000 Ω to 5200 Ω: 200 W
At 10 % duty cycle (5 seconds on, 45 seconds off):
10 Ω: 300 W
20 Ω to 2400 Ω: 500 W
2425 Ω to 2900 Ω: 400 W
3000 Ω to 5200 Ω: 200 W
Current
RMS: 0 mA to 5,500 mA
Accuracy:
< 50 mA: +/- (3.5 % + 1 mA)
≥ 50 mA: +/- (2.5 % + 1 mA)
Voltage
Peak: 10 kV Peak to Peak
Accuracy: ± (10 % of reading + 50 V)
Crest factor: 1.4 to 16.0
Defined as ratio of Peak voltage to RMS voltage (Vpk / Vrms), using the larger peak
Vessel Sealing Measurement
Loop current, RMS: 0 mA to 5500 mA
Accuracy:
< 50 mA: +/- (3.5 % + 1 mA)
≥ 50 mA: +/- (2.5 % + 1 mA)
HF Leakage Current
Fixed load: 200 Ω
Voltage accuracy: ±2.5 %
Power rating: 400 W
Additional fixed load 200 Ω
Current, RMS: 0 mA to 5500 mA
Accuracy:
< 50 mA: +/- (3.5 % + 1 mA)
≥ 50 mA: +/- (2.5 % + 1 mA)
CQM Test (Contact Quality Monitor)
Resistances: 0 Ω to 475 Ω (by 1 Ω)
Accuracy: 0 Ω to 10 Ω ±0.5 Ω, 11 Ω and above ±5 %
Power rating: 0.5 W
Auto time interval: 1 to 5 seconds
Oscilloscope Output
1 V per ampere of input current, typical
Footswitch Simulations
Cut and Coag
Communications
USB device port: Micro B connector, full speed
Memory
Test records: 5,000
Non-volatile retained through power cycling
Calibration: Recommended cycle: Traceable to the International System of Units (SI) through appropriate National Metrology Institutes such as NIST or intrinsic standards