Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Flow rate measurement
Flow is calculated by measuring volume over time
- Range: 0.5 ml/h to 1000 ml/h
- Accuracy: 1 % of reading ± 1 LSD for flows of 16 ml/h to 200 ml/h for volumes over 20 ml, otherwise 2 % of reading ± 1 LSD for volumes over 10 ml under laboratory conditions.
- Max test duration: 10 hours on battery
Volume measurement
Volume measured directly by the measuring module in minimum sample sizes of 60 microliters
- Range: 0.06 ml to 999 ml
- Accuracy: 1 % of reading ± 1 LSD for flow rates of 16 ml/h to 200 ml/h for volumes over 20 ml. Otherwise 2 % of reading ± 1 LSD for volumes over 10 ml under laboratory conditions.
- Max test duration:10 hours on battery
Pressure measurement
Direct measurement of pressure at the inlet port
- Range: 0 psi to 45 psi or equivalent in mmHg and kPa
- Accuracy: 1 % of full scale ± 1 LSD under laboratory conditions
- Max test duration: 30 minutes
General specifications
Storage of Results
Test results stored for later viewing, printing or transfer to PC. Typical practical capacity: 100 tests
Power Down
The results of tests in progress will be saved in the case of accidental power down
Computer Control
The Product can be fully controlled from a PC using HydroGraph V3 software for IDA-1S
Battery Power: 4 x Panasonic HHR210AB NiMh 2000 mAh batteries
Dimensions: 30 cm x17 cm x 10 cm (12 in x 8 in x4 in)
Weight: ~1.2 kg (2.7 lbs)
Safety
IEC 61010-1: Overvoltage category II, Pollution Degree 2
Electromagnetic Environment
IEC 61326-1: Basic
Emissions Classification
IEC CISPR 11: Group 1, Class A.