Calibration Range: 100-2500μg/m3
Measuring Range: 100-7000μg/m3
Sampling (flow) Rate: 200mL/min
Sampling Volume: 6000mL, 2000mL
Sampling Time: 30 min, 10 min
Colour Change: White → Pale brown
Nr. of tests: 5 tubes
Automatic Gas Sampler: GSP-300FT-2
Catalog
Specifications
| Calibration Range |
100-2500μg/m3 |
| Measuring range |
100-7000μg/m3 |
| Sampling (flow) Rate |
200mL/min |
| Sampling Volume |
6000mL, 2000mL |
| Sampling Time |
30min, 10min |
| Detecting Limit |
50μg/m3(6000mL) |
| Colour Change |
White → Pale brown |
| Corrections for temperature/humidity |
Unnecessary for 5 to 35 ºC Unnecessary for R.H.0 to 80% |
| Shelf life |
2 years |
Possibly coexisting substances and their interferences
(NOTE)
| Substance(s) |
Interference |
Colour change by itself |
| Ethanol |
|
No |
No |
| Xylene,Styrene |
1000μg/m3 |
+ |
Pale brown |
| p-Dichlorobenzene |
|
No |
No |
| α-Pinene |
|
No |
No |
| n-Hexane |
|
No |
No |
| Aromatic hydrocarbons |
|
+ |
Pale brown |
| Formaldehyde |
1/3 |
No |
No |
Other substance(s) measurable with this detector tube
| Substance(s) |
Correction Factor |
Pump Strokes [n] |
Measuring Range |
| Ethyl benzene |
coefficient: 1.1 |
200mL/min x 30min |
110-2750μg/m3 |
| Xylene |
coefficient: 5.4 |
200mL/min x 30min |
540-13500μg/m3 |
| Calibration gas generation |
Permeation tube method |
Miscellaneous
Electric sampling pump This detector tube is suited for measuring toluene in indoor environments. The GSP-200 / GSP-300FT-2 series automatic gas samplers (or comparable sampling devices/pumps) should always be used for optimal performance. A deviation of +/-10% in the sampling flowrate (i.e. amount of air flowing through the tube at a given moment) or the sampling amount (total sampled amount of target substance) causes an average margin of error of approx. +/-10% (maximum +/-20%) in the indicated value. |
| TLV-TWA |
TLV-STEL |
Explosive range |
| 20 ppm(2014) |
-- |
1.4。チ6.7% |