Indicator: 0.05 ~ 0.6%
Measuring Range: 0.05 ~ 0.6%, Number of Pump Strokes 2
Sampling Time: 45 seconds per pump stroke
Colour Change: Orange →Greenish brown
Reaction Principle: (CH3)2CHCH2COCH3+ Cr6+ + H2SO4 → Cr3+
Coefficient of Variation: 10% (for 0.05 to 0.6 %)
Nr. of tests: 10
Pump: GV-100S/110S
Catalog
Manual
Performance
| Measuring Range |
0.05 to 0.6% |
| Number of Pump Strokes |
2 |
| Correction Factor |
1 |
| Sampling Time |
45 seconds per pump stroke |
| Detecting Limit |
0.005% (n=2) |
| Colour Change |
Orange →Greenish brown |
| Reaction Principle |
(CH3)2CHCH2COCH3+ Cr6+ + H2SO4 → Cr3+ |
| Coefficient of Variation |
10% (for 0.05 to 0.6 %) |
| Shelf Life |
3 Years |
| Corrections for temperature & humidity |
Temperature correction is necessary |
| Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
| Substance |
Concentration |
Interference |
Change colour by itself to |
| Sulphur dioxide |
100 ppm |
Two layers (Greenish brown & Pale brown) |
Brown( 50 ppm) |
| Hydrogen sulphide |
100 ppm |
Two layers (Greenish brown & Pale brown) |
Greenish brown( 25 ppm) |
| Alcohols, Esters,Ketones |
|
+ |
Greenish brown |
| Propane |
0.2% |
Cannot use |
Pale brown for whole layer( 0.2%) |
| Toluene |
25ppm |
Two layers (Greenish brown & Pale brown) |
Pale brown( 15 ppm) |
Other substance measurable with this detector tube
| Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
| Styrene |
by scale |
2 |
0.15 to 2.3% |
| Calibration gas generation |
Static gas dilution method |
| TLV-TWA |
TLV-STEL |
Explosive range |
| 20 ppm (2015) |
75 ppm (2015) |
1.2 to 8% |