Indicator: 10 ~ 50 mg/L
Measuring Range: 5 ~ 50 mg/L
Sampling Time: 5 minute
Colour Change: White → Orange
Reaction Principle: Fe2+ + 1,10 Fenantoroline (White) → Complex compound
Nr. of tests: 10
Catalog
Performance
| Measuring Range |
(5) to 50 mg/L |
| Sampling Time |
5 minutes |
| Detecting Limit |
1.0 mg/L |
| Colour Change |
White → Orange |
| Reaction Principle |
Fe2+ + 1,10 Fenantoroline (White) → Complex compound |
| Shelf Life |
3 Years |
| Corrections for temperature & humidity |
Unnecessary |
| Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
| Substance |
Concentration |
Interference |
Change colour by itself to |
| Zn |
≧ 5 mg/L |
+ |
No ( ≦ 100mg/L) |
| Co |
≧ 5 mg/L |
+ |
Orange ( ≧ 1mg/L) |
| CN- |
≧ 20 mg/ L |
 |
No ( ≦ 100mg/L) |
| Fe3+ |
≧ 50 mg/L |
+ |
Pale orange ( ≧ 50 mg/L) |
| CU+ |
≧ 5 mg/L |
 |
Pale orange (≧ 1 mg/L) |
| CU2+ |
≧ 10 mg/L |
+ |
Pale blue ( ≧ 50 mg/L) |
| Ni |
≧ 5 mg/L |
+ |
No ( ≦ 100 mg/L) |
| Mn |
≧ 50 mg/L |
 |
No ( ≦ 100 mg/L ) |
| PO43- |
≦ 100 mg/L |
No |
No ( ≦ 100 mg/L) |
Detector tubes for dissolved substances
(i.e. in solution).
Dissolved sulphide detector tube
211H / 211M / 211 / 211LL

|
 |
Liquid detector tube
| Ozone: |
218 |
| Chloride ion: |
221L / 221LL |
| Mercury: |
271 |
| Hexavalent chromium: |
273 |
| Bivalent iron: |
281 |
| Copper: |
284 |
| Zinc: |
285 |
| Nicke: |
291 |
|
To measure, simply immerse the detector tube in a sample of liquid (placed in a beaker or other container). Anyone can measure concentrations quickly, easily and accurately on site.
* Adjustment for pH is required when measuring with liquid detector tubes.