Indicator: 0.05-1ppm
Measuring Range: 0.05-1ppm, Number of Pump Strokes 5
Sampling Time: 1.5 minute per pump stroke
Colour Change: Yellowish Brown → Reddish Brown
Reaction Principle: 3HCHO + (NH2OH)3H3PO4 → H3PO4, H3PO4 + Base → Phosphate
Coefficient of Variation: 10% (for 0.05 to 0.3 ppm), 5% (for 0.5 to 1.0 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Catalog
Manual
manual tube
Performance
| Measuring Range |
0.05 to 1.0 ppm |
| Number of Pump Strokes |
5 |
| Correction Factor |
1 |
| Sampling Time |
1.5 minutes per pump stroke |
| Detecting Limit |
0.03 ppm (n=5) |
| Colour Change |
Yellowish Brown → Reddish Brown |
| Reaction Principle |
3HCHO + (NH2OH)3H3PO4 → H3PO4 H3PO4 + Base → Phosphate |
| Coefficient of Variation |
10% (for 0.05 to 0.3 ppm), 5% (for 0.5 to 1.0 ppm) |
| Shelf Life |
1 Year |
| Corrections for temperature & humidity |
Temperature correction is necessary |
| Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Possible coexisting substances and their interferences
(NOTE)
| Substance |
Concentration |
Interference |
Change colour by itself to |
| Ammonia |
|
 |
Yellow (Unclear demarcation) |
| Nitrogen dioxide |
|
No |
Pale pink |
| Aldehydes |
1/3 time |
+ (Unclear demarcation) |
Reddish brown |
| Ketones |
1:1 |
+ (Unclear demarcation) |
Reddish brown |
| Toluene |
10 ppm |
No |
No |
| Calibration gas generation |
Diffusion tube method |
| TLV-TWA |
TLV-STEL |
Explosive range |
| - |
C 0.3 ppm |
7.0 to 73% |