GENERAL SPECIFICATIONS
OPERATING AMBIENT CONDITIONS
TEMPERATURE: 18 °C to 28 °C
HUMIDITY: Up to 80 % RH
STORAGE CONDITIONS
TEMPERATURE : -20 °C to 50 °C
HUMIDITY: 0 % to 95 % RH (non-condensing)
ALTITUDE : 2000 m
CALIBRATION SYSTEM ACCURACY
TEMPERATURE: ± 0.10 °C Or Better Typically ± 0.05 °C Based on Probe Accuracy
HUMIDITY : ± 1.00 % RH Or Better Specially tuned systems can be as good as ± 0.60 % Based on Probe Accuracy
CONTROLLER TYPE : PID Controller
UNIFORMITY
TEMPERATURE : 0.10 °C Or Better Typically ± 0.05 °C
HUMIDITY: 0.30 % RH @ 18 °C to 28 °C Or Better Typically ± 0.25 %
STABILITY
TEMPERATURE : 0.05 °C
HUMIDITY : 0.15 % RH @ 18 °C to 28 °C
WARM-UP PERIOD
AMBIENT CONDITIONS : 30 Minutes Maximum
COLD :30 Minutes
RESOLUTION
DISPLAY
Temperature : 0.01 °C
Humidity: 0.01 % RH
Dew Point : 0.01 °C (Calculated)
USB
Temperature 0.01 °C
Humidity 0.01 % RH
RAMP / SOAK RATE OF CHANGE
Depends on ambient and desiccant conditions
TEMPERATURE
High → Low 1.50 °C / minute Typical
Low → High 5.00 °C / minute Typical
HUMIDITY
High → Low 5.00 % RH / minute Typical
Low → High10.00 % RH / minute Typical
CONTROL RANGE
TEMPERATURE 5 °C to 55 °C
HUMIDITY@ 18°C 5 % to 95 % RH HUMIDITY@ 35°C 5 % to 80 % RH
HUMIDITY@ 23°C 5 % to 95 % RH HUMIDITY@ 40°C 5 % to 75 % RH
HUMIDITY@ 30°C 5 % to 85 % RH HUMIDITY@ 50°C 5 % to 60 % RH