CONTROL
Range : RH 5% to 95% @18°C-30°C
Temperature 5 to 55°C
Uniformity : RH ±0.2% @5% to 50%
RH ±0.3% @50% to 90%
RH ±0.3% @90% to 95%
Temperature ±0.2°C @50%RH
Stability at Equilibrium: RH ± 0.3% @ 23°C
Temperature ±0.1°C
Controller Type : Embedded multiple input PID controller, LCD
Resolution
RH 0.1%
Temperature 0.1°C
Generation Method: Mixed flow with desiccant dryer cell and heating element humidifier
PERFORMANCE
Response Time:
RH 3-5 Minutes
Temperature Cooling : 1.5°C/Minute
Heating : 5°C/Minute
Calibration Uncertainties
RH ± 0.7% @5-80%
RH ± 1.0% @80-95%
Temperature : ± 0.2°C
Control Probe : RH Uncertainty ±0.5% @23°C