GENERAL SPECIFICATIONS
OPERATING AMBIENT CONDITIONS
TEMPERATURE :18 °C to 28 °C
HUMIDITY: Up to 80 % RH
STORAGE CONDITIONS
TEMPERATURE: -20 °C to 50 °C
HUMIDITY: 0 % to 95 % RH (non-condensing)
ALTITUDE: 2000 m
CONTROLLED RANGE
TEMPERATURE : 5 °C to 55 °C
HUMIDITY@ 18°C 5 % to 95 % RH HUMIDITY@ 35°C 5 % to 80 % RH
HUMIDITY@ 23°C 5 % to 95 % RH HUMIDITY@ 40°C 5 % to 75 % RH
HUMIDITY@ 30°C 5 % to 85 % RH HUMIDITY@ 50°C 5 % to 60 % RH
CALIBRATION SYSTEM ACCURACY
TEMPERATURE: ± 0.10 °C Or Better Typically ± 0.05 °C
HUMIDITY: ± 1.00 % RH Or Better Specially tuned systems can be as good as ± 0.60 %
CONTROLLER TYPE: PID Controller
UNIFORMITY
TEMPERATURE : 0.10 °C Or Better Typically ± 0.05 °C
HUMIDITY : 0.30 % RH @ 18 °C to 28 °C Or Better Typically ± 0.25 %
STABILITY
TEMPERATURE: 0.05 °C
HUMIDITY : 0.15 % RH @ 18 °C to 28 °C
RAMP / SOAK RATE OF CHANGE
Depends on ambient conditions
Temperature
High → Low 1.50 °C / minute (Typical)
Low → High 5.00 °C / minute (Typical)
Humidity
High → Low 5.00 % RH / minute (Typical)
Low → High 10.00 % RH / minute (Typical)
WARM-UP PERIOD
Ambient Condition : 30 Minutes Maximum
Cold : 30 Minutes
RESOLUTION
DISPLAY
Temperature 0.01 °C
Humidity 0.01 % RH
Dew Point 0.01 °C (Calculated)
USB
Temperature 0.01 °C
Humidity 0.01 % RH