CALIBRATION
Traceability 17025 Traceable to NIST
OPERATING AMBIENT CONDITIONS
Temperature: 18 °C to 28 °C
Humidity: Up to 80 % RH
CONTROLLED RANGE
Temperature: 5 °C to 55 °C
Humidity @18 °C 5 % to 95 % RH
Humidity @30 °C 5 % to 75 % RH
Humidity @23 °C 5 % to 95 % RH
Humidity @40 °C 5 % to 60 % RH
Humidity @28 °C 5 % to 80 % RH
Humidity @50 °C 5 % to 40 % RH
CALIBRATION SYSTEM ACCURACY
Temperature : +/- 0.10 °C
Humidity: +/- 1.00 % RH Or Better Typically +/- 0.50 %
Controller type : PID Controller
UNIFORMITY
Temperature: 0.10 °C
Humidity: 0.30 % RH @ 23 °C
STABILITY
Temperature : 0.05 °C
Humidity 0.15 % RH @ 23 °C
RESOLUTION DISPLAY
Temperature: 0.01 °C
Humidity: 0.01 % RH
Dew Point : 0.01 °C (Calculated)
RAMP / SOAK RATE OF CHANGE
Temperature
High → Low 0.50 °C / minute Typical
Low → High 1.00 °C / minute Typical
Humidity
High → Low 1.00 % RH / minute Typical
Low → High 5.00 % RH / minute Typical