Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

CALIBRATION TO (17025 TRACEABLE TO NIST)
Transfer standard: No Charge
Certificate: 17025 Validation (Additional Fees Apply)
OPERATING AMBIENT CONDITIONS
Temperature : 18 °C to 28 °C
Humidity : Up to 80 % RH
STORAGE CONDITIONS
Temperature: -20 °C to 50 °C
Humidity: 0 % to 95 % RH (non-condensing)
Altitude: 2000 m
CALIBRATION SYSTEM ACCURACY
Temperature: ± 0.10 °C Or Better Typically ± 0.05 °C *Based on Probe Accuracy
Humidity: ± 1.00 % RH Or Better Specially tuned systems can be as good as ± 0.60 % *Based on Probe Accuracy
Controller type: PID Controller
UNIFORMITY
Temperature :0.10 °C Or Better Typically ± 0.05 °C
Humidity: 0.30 % RH @ 18 °C to 28 °C Or Better Typically ± 0.25 %
STABILITY
Temperature : 0.05 °C
Humidity: 0.15 % RH @ 18 °C to 28 °C
WARM-UP PERIOD
Ambient conditions : 30 Minutes Maximum
Cold : 30 Minutes
RESOLUTION
Display
Temperature 0.01 °C
Humidity 0.01 % RH
Dew Point 0.01 °C (Calculated)
USB
Temperature 0.01 °C
Humidity 0.01 % RH
RAMP / SOAK RATE OF CHANGE
Depends on ambient and desiccant conditions
Temperature
High → Low 1.50 °C / minute Typical
Low → High 5.00 °C / minute Typical
Humidity
High → Low 5.00 % RH / minute Typical
Low → High 10.00 % RH / minute Typical
CONTROL RANGE
Temperature 5°C to 55 °C
Humidity @ 18°C 5 % to 95 % RH
Humidity @ 35°C 5 % to 80 % RH
Humidity @ 23°C 5 % to 95 % RH
Humidity @ 40°C 5 % to 75 % RH
Humidity @ 30°C 5 % to 85 % RH
Humidity@ 50°C 5 % to 60 % RH