GENERAL SPECIFICATIONS (For Each Chamber - Chamber 1,2 and 3)
OPERATING AMBIENT CONDITIONS
TEMPERATURE : 18 °C to 28 °C
HUMIDITY: Up to 80 % RH
STORAGE CONDITIONS
TEMPERATURE : -20 °C to 50 °C
HUMIDITY: 0 % to 95 % RH (non-condensing)
ALTITUDE: 2000 m
CALIBRATION SYSTEM ACCURACY
TEMPERATURE : ± 0.12 °C Or Better Typically ± 0.10 °C Based on Probe Accuracy
HUMIDITY : ± 1.00 % RH Or Better Typlically ± 0.80 % Based on Probe Accuracy
CONTROLLER TYPE : PID Controller
UNIFORMITY
TEMPERATURE : 0.10 °C Or Better Typically ± 0.10 °C
HUMIDITY : 0.30 % RH @ 18 °C to 28 °C Or Better Typically ± 0.25 %
STABILITY
TEMPERATURE: 0.05 °C
HUMIDITY: 0.15 % RH @ 18 °C to 28 °C
WARM-UP PERIOD
AMBIENT CONDITIONS : 30 Minutes Maximum
COLD : 30 Minutes
RESOLUTION
DISPLAY
Temperature: 0.01 °C
Humidity : 0.01 % RH
Dew Point : 0.01 °C (Calculated)
USB
Temperature: 0.01 °C
Humidity: 0.01 % RH
RAMP / SOAK RATE OF CHANGE
Depends on ambient and desiccant conditions
TEMPERATURE
High → Low 1.50 °C / minute *Typical
Low → High 3.00 °C / minute *Typical
HUMIDITY
High → Low 5.00 % RH / minute *Typical
Low → High 10.00 % RH / minute *Typical