Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Range:
- Sensor: 0-50 NTU and 0-1000 NTU
- Meter: 0-50 NTU or 0-1000 NTU selectable
Accuracy: + 1% of full scale
Meter Resolution: 12 bit
Output:
- Sensor: 4-20mA
- Meter: LED screen
Method: Nephelometer with correction
Operating Voltage:
- Sensor: 10-36 VDC @ 40 MS
- Meter: Internal 9VDC battery
Current Draw: 30 mA plus sensor output (Sensor)
Warm-up Time: 5 seconds minimum (Sensor)
Operating Temperature:
- Sensor: 14 to 122°F (-10 to +50°C)
- Meter: 32 to 122°F (0 to +50°C)
Materials: 306 stainless steel, delrin, polyether jacketed cable.
Maximum Pressure: 30 psi
Light Source: Infrared LED, (880nm)
Sensor Cable Length: 25 ft standard (optional to 500 ft)
Size: Body= 1 1/2 x 8.5 inches (3.8 x 21.6 cm) (Dia x Length)
Weight: 2 lbs (907 g) (Meter+sensor)