Booster capacity: 1.5 KVA
Step up speed: 0.5kv/s-5.0kv/s (increasing every 0.5KV) 10 scales, error 0.2kv/s
Output voltage: 0-80 kV
Voltage accuracy: (2% reading + 2 digits)
Power distortion rate: < 1%
Electrode gap: standard 2.5mm
Test times: 6 times (1-9 times for optional)
Static release time: 900s (0-9000s for optional)
Interval rest time: 300s (0-900s for optional)
Mixing time: 15s (0-250s for optional)
Dimension: 650mm × 470mm × 410mm
Weight: 35kg
Operating Conditions
Ambient temperature: 0-40 ℃
Relative humidity: ≤ 85%
Working power: AC 220V (1 ± 10%)
Power frequency: 50/60 Hz
Power consumption: < 200 W
Accessories
Oil cup & Electrode: 1 set
Mushroom shape electrode head: 1 pair
Spherical shape electrode head: 1 pair
Flat shape electrode head: 1 pair
Power supply cable: 1pc
Standard gauge: 1 pc
Magnon: 2pcs
Tweezer: 1pc
Earth wire: 1 pc
Fuse (3A): 2 pcs
Print paper: 1pc
Operating manual: 1pc
Instrument features
1) The instrument is controlled by large capacity single chip microcomputer, which is stable and reliable;
2) The instrument has temperature and humidity and clock display functions, and can customize infrared oil temperature measurement.
3) Built-in wide range watchdog circuit inside of the instrument to eliminate the phenomenon of crash;
4) Variety of standard selection, the instrument program is equipped with GB / T507-1986, GB / T507-2002, DL429.9, IEC156 and self- programming operation, which can adapt to various choices of different users;
5) Oil cup of the instrument is made of special glass and polymer materials, which is finely processed to prevent oil leakage and corrosion;
6) Unique high-voltage sampling design of the instrument allows the test value to enter the A / D converter directly, which avoids the error caused in the analog circuit and makes the measurement result more accurate;
7) The instrument has over-current, over-voltage, short-circuit protection and other functions, and has strong anti-interference ability and good electromagnetic compatibility;
8) Can switch between Chinese and English at will to meet different customer requirements;
9) The instrument has the function of data analysis to assist the staff to determine oil quality;
10) USB or RS232 data transmission;
11) Large wide colour touch screen display.