- Frequency bandwidth: DC to 100 MHz (-3 dB)
- Rise time: 3.5 ns or shorter
- Noise level: 2.5 mA rms max. (bandwidth limited to 20 MHz)
- Continuous allowable input: 30 A rms (requires derating at frequency)
- Max. allowable peak input: 50 A peak (non continuous)
- Amplitude accuracy: (30 min. after power-on, after degaussing and zero-adjustment) ±1.0 % rdg. ±1 mV f.s. (DC, 45 to 66 Hz, 0 to 30 A rms)
±2 % rdg. (DC, 45 to 66 Hz, 30 A rms to 50 A peak)
- Output rate: 0.1 V/A *Device output is terminated internally. Use with a device having a 1 MΩ input resistance or higher.
- Measurable conductors: Insulated conductor
- Core diameter: φ 5 mm (0.20 in)
- Power supply :±12 V ±0.5 V, 5.3 VA max.
- Dimensions and mass: 175 mm (6.89 in)W × 18 mm (0.71 in)H × 40 mm (1.57 in)D, 240 g (8.5 oz)
- Sensor cable BNC terminal: 1.5 m (4.92 ft), Power cable: 1 m (3.28 ft)
- Accessories: Instruction manual ×1, Carrying case × 1