- Rated primary current: AC 10 A (for leak current measurement, 50/60 Hz)
- Max. allowable input (45-66 Hz): Max. 10 A of continuous input
- Rated output current: AC 100 mV/A
- Amplitude accuracy (45-66 Hz): ±1.0 % rdg. ±0.005 % f.s.
- Phase accuracy: ±5 ° (at 50 Hz or 60 Hz)
- Frequency characteristics: 40 Hz to 5 kHz: ± 5%
(Add to amplitude accuracy)
- Max. rated voltage to earth: 40 Hz to 5 kHz: ± 5%
(Add to amplitude accuracy)
- Core diameter: φ 30 mm (1.18 in)
- Dimensions and mass: 60 mm (2.36 in)W × 112.5 mm (4.43 in)H × 23.6 mm (0.93 in)D, 160 g (5.6 oz), Cord length: 3 m (9.84 ft)
- Accessories: Instruction manual ×1
- Rated primary current: AC 10 A (for leak current measurement, 50/60 Hz)
- Max. allowable input (45-66 Hz): Max. 10 A of continuous input
- Rated output current: AC 100 mV/A
- Amplitude accuracy (45-66 Hz): ±1.0 % rdg. ±0.005 % f.s.
- Phase accuracy: ±5 ° (at 50 Hz or 60 Hz)
- Frequency characteristics: 40 Hz to 5 kHz: ± 5%
(Add to amplitude accuracy)
- Max. rated voltage to earth: 40 Hz to 5 kHz: ± 5%
(Add to amplitude accuracy)
- Core diameter: φ 30 mm (1.18 in)
- Dimensions and mass: 60 mm (2.36 in)W × 112.5 mm (4.43 in)H × 23.6 mm (0.93 in)D, 160 g (5.6 oz), Cord length: 3 m (9.84 ft)
- Accessories: Instruction manual ×1