Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

- Rated input current: 500A AC/DC
- Max. allowable input: 700 Arms (1000 Apeak, requires derating at frequency)
- Frequency characteristics: Amplitude: DC to 100 kHz
- Phase: DC to 100 kHz
- Amplitude and Phase accuracy: DC, 45 Hz ≤ f ≤ 66 Hz
(±0.05 % rdg. ±0.01 % f.s. (Phase: ±0.2 deg) Defined to 100 kHz
- Power consumption: 5 VA max. (at 500 A/55 Hz, ±12 V power requirement)
- Output voltage rate: 2 V /rated current value (voltage output with the Sensor Unit 9555-10, use with a device having a 1 MΩ input resistance or higher)
- Max. rated voltage to earth: AC/DC 1000 V (50/60 Hz, CAT III )
- Core diameter: φ 36 mm (1.42 in)
- Operating temperature, humidity: 0°C to +50°C (32°F to 122°F)
80% rh or less (with no condensation)
- Power supply : DC ±11 V to ±15 V (Power suppled via the 9555-10, which supports 100 to 240 V AC)
- Dimensions and mass: 160 mm (6.30 in)W × 112 mm (4.41 in)H × 50 mm (1.97 in)D, 850 g (30.0 oz), cord length: 3 m (9.84 ft)
- Accessories: Instruction manual ×1, Mark bands ×6
- Rated input current: 500A AC/DC
- Max. allowable input: 700 Arms (1000 Apeak, requires derating at frequency)
- Frequency characteristics: Amplitude: DC to 100 kHz
- Phase: DC to 100 kHz
- Amplitude and Phase accuracy: DC, 45 Hz ≤ f ≤ 66 Hz
(±0.05 % rdg. ±0.01 % f.s. (Phase: ±0.2 deg) Defined to 100 kHz
- Power consumption: 5 VA max. (at 500 A/55 Hz, ±12 V power requirement)
- Output voltage rate: 2 V /rated current value (voltage output with the Sensor Unit 9555-10, use with a device having a 1 MΩ input resistance or higher)
- Max. rated voltage to earth: AC/DC 1000 V (50/60 Hz, CAT III )
- Core diameter: φ 36 mm (1.42 in)
- Operating temperature, humidity: 0°C to +50°C (32°F to 122°F)
80% rh or less (with no condensation)
- Power supply : DC ±11 V to ±15 V (Power suppled via the 9555-10, which supports 100 to 240 V AC)
- Dimensions and mass: 160 mm (6.30 in)W × 112 mm (4.41 in)H × 50 mm (1.97 in)D, 850 g (30.0 oz), cord length: 3 m (9.84 ft)
- Accessories: Instruction manual ×1, Mark bands ×6








