Measurement line type :Isolated input, DCCT input
Rated primary current: 50 A AC/DC
Measurement line type :Isolated input, DCCT input
Rated primary current: 50 A AC/DC
Number of input channels: 3 channels
Maximum input current: 60 A
Amplitude and Phase accuracy: DC (±0.02 % rdg. ±0.007 % f.s.)
45 Hz < f ≤ 65 Hz (±0.02 % rdg. ±0.005 % f.s., Phase: ±0.1 deg.)
Accuracy is defined to 1 MHz
Output voltage: 2 V/50 A
Measurement terminals: Terminal block (with safety cover), M6 screws
Input resistance: 1.5 mΩ or less (50 Hz/60 Hz)
Input capacitance: Between measurement terminals and case (secondary side), 40 pF or less, defined at 100 kHz
Power supply: Power supply from PW6001, 3390, 3390-10
Dimensions and mass 430 mm (16.93 in) W × 88 mm (3.46 in) H × 260 mm (10.24 in) D, Cable length: 0.8 m (2.62 ft)
PW9100-03: 3.7 kg (130.5 oz), PW9100-04: 4.3 kg (151.7 oz)
Accessories :Instruction Manual ×1