Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

-Conductivity (EC) Range:
0.000 µS/cm to 20.00 mS/cm (k=0.1)
0.00 µS/cm to 200.0 mS/cm (k=1)
0.0 µS/cm to 2.000 S/cm (k=10)
Resolution: 0.05% of full scale
Accuracy: ± 0.6% of full scale; ± 1.5% of full scale > 18.0 mS/cm
Reference Temperature: 15 to 30 °C (Adjustable)
Temperature Coefficient: 0.00 to 10.00 % (Adjustable)
Cell Constant: 0.1 / 1.0 / 10.0 (Selectable)
Measurement Unit: μS/cm, mS/cm, S/m (Auto-ranging / Manual)
Calibration Points: Up to 4 (Auto / Manual)
TDS Range: 0.00 mg/L to 100 g/L (TDS factor=0.40 to 1.00)
Resolution: 0.01 mg/L
Accuracy: ± 0.1% of full scale
Calibration Curves: EN27888 / 442 / Linear / NaCl
-Resistivity Range:
0.0 Ω•cm to 2.000 MΩ•cm (k=10)
0.000 kΩ•cm to 20.00 MΩ•cm (k=1)
0.00 kΩ•cm to 200.0 MΩ•cm (k=0.1)
Resolution: 0.05% of full scale
Accuracy: ± 0.6% of full scale; ± 1.5% of full scale > 1.80 MΩ•cm
Temperature Range: -30.0 to 130.0 °C
Resolution: 0.1 °C
Accuracy: ± 0.4 °C
Calibration Option: Yes
Memory Data: 100
Auto-Shut Off: Yes (Programmable: 1 to 30 minutes)
Auto-Hold / Auto-Stable: Yes
Cell Constant Display: Yes
Electrode Status: On screen display
Diagnostic Messages: Yes