Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

(Full specifications can be found on datasheet)
Linearity better than 20 ppm (5-dial model)
Input impedance: for 1455-A 10 kΩ & 1455-B, 1455-BH 100 kΩ
High-frequency model, down 3 dB at 7.5 MHz
BASIC SPECIFICATIONS
|
1455-A
|
1455-B
|
1455-BH
|
|
| Number of Dials: |
4
|
5
|
5
|
| Input Resistance: |
10 kΩ
|
10 kΩ
|
100 kΩ
|
| Accuracy of Input R: (ppm) |
150
|
150
|
150
|
| Input Voltage Rating: |
230 V
|
230 V
|
700 V
|
| Frequency Response: |
850 kHz
|
690 kHz
|
69 kHz
|
| Resolution: (ppm of input) |
100
|
10
|
10
|
| Linearity (sum of A & B) | |||
| A. Absolute Linearity | |||
| 0.00001 to 0.00010 |
n/a
|
0.03
|
0.02
|
| 0.00010 to 0.00100 |
0.3
|
0.3
|
0.2
|
| 0.00100 to 0.01000 |
2
|
3
|
2
|
| 0.01000 to 0.10000 |
15
|
10
|
10
|
| 0.10000 to 1.00000 |
30
|
20
|
20
|
| B. Terminal Linearity (ppm of input). | |||
| FOUR-TERMINAL (low output terminal) |
0.04
|
0.04
|
0.004
|
| THREE-TERMINAL |
0.2
|
0.2
|
0.02
|
The GR 1455 Kelvin-Varley Voltage Dividers provide accurately known voltage ratios from 0.00001 to 1.00000 for use in many common measurements:
A resistive divider of the Kelvin-Varley type, the 1455 has precision resistors throughout (rather than in selected positions only) for over-all high accuracy. Linearity is as low as 0.02 ppm of input.