Min. detectable leak rate for helium (vacuum method): <5∙10-12 mbar∙L/s
Min. detectable leak rate for helium (sniffer mode): <5∙10-8 mbar∙L/s
Max. inlet pressure (MASSIVE mode): atm. (~1013 mbar)
Type of roughing pump: roots
Max. pumping speed of roughing pump: 36 m³/h
Pumping speed @ inlet (100-1000 mbar): 23 m³/h
Pumping speed @ inlet (10-100 mbar): 26 m³/h
Pumping speed @ inlet (1-10 mbar): 26 m³/h
Max. helium pumping speed (ULTRA mode):: >36 L/s
Max. helium pumping speed DN63 reduced to DN40: ~ 30 L/s
Response time for 50 L: 1.4 s
Evacuation time to 1 mbar 50 L: 52 s
Evacuation time to 1 mbar 100 L*: 104 s
Evacuation time to 1 mbar 500 L*: 527 s
Evacuation time to 1 mbar 1000 L*: 1054 s
Venting 50 L to atm.: 16 s
Run up time for operation: <2 min
Time for calibration: <30 s
Time for rate-of-rise measurement (I·RISE): <10 s
Inlet flange: DN63 ISO-K
Detectable masses (ULTRATEST Sensor Technology): 2, 3, 4 (H2, 3He, He)
Supply voltage: 100 bis 230 V 50/60 Hz
Max. Power Consumption: 1500 VA
Typ. Power Consumption: 700 VA
Dimensions (l x w x h): 1050 x 472 x 1040 mm
Weight: 144 kg