JEIOtech LBV-040 Heating chamber
Controller: Programmable, Touch screen type controller (3.5” color monitor)
Interface port: Basically RS-232C [RS-485 (maximum 32 units control) -optional]
Air flow velocity (m/s): 0.2 to 0.6
Chamber volume (L / cu ft): 420 / 14.8
Air change rate (approx, x/h): 173
Air circulation (approx, x/h): 294
Max. load per shelf (kg / lbs): 30 / 66.1
Permitted total load (kg / lbs): 90 / 198.4
Electrical requirements (380V, 50Hz): 3ph / 9.4A
No. of wire shelves (standard / max.): 2 / 22
Distance of between shelves (mm / inch): 30 / 1.2
Exhaust air volume flow (approx, L/min): 1400
Dimension (W×D×H) - Exterior (mm / inch): 1350×1045×1712/ 53.1×41.1×67.4
Dimension (W×D×H) - Interior (mm / inch): 750×700×800/ 29.5×27.6×31.5
Heating time 2) (min.) from Amb. to 250℃: 51
Dimension (W×D×H) - Net weight (kg / lbs): 320 / 705.5
Temperature fluctuation 1) at 100℃ (±℃ / ℉): 0.3 / 0.54
Temperature variation 2) (±℃ / ℉) - at 100℃: 1.3 / 2.34
Temperature variation 2) (±℃ / ℉) - at 150℃: 2.3 / 4.14
Temperature variation 2) (±℃ / ℉) - at 200℃: 3.2 / 5.76
Temperature variation 2) (±℃ / ℉) - at 250℃: 4.1 / 7.38
Temperature range (℃ / ℉) - intake/damper 100% open: Max. 120 / Max. 248
Temperature range (℃ / ℉) - intake/damper 100% close: Amb.+45 to 250 / Amb. +81 to 482
Recovery time 2) after door was opened (min.) - at 100℃: 8
Recovery time 2) after door was opened (min.) - at 150℃: 9
Recovery time 2) after door was opened (min.) - at 200℃: 8.5
Recovery time 2) after door was opened (min.) - at 250℃: 9.5
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

