JEIOtech LBV-070 Heating chamber
Controller: Programmable, Touch screen type controller (3.5” color monitor)
Interface port: Basically RS-232C [RS-485 (maximum 32 units control) -optional]
Air flow velocity (m/s): 0.2 to 0.6
Chamber volume (L / cu ft): 720 / 25.4
Air change rate (approx, x/h): 233
Air circulation (approx, x/h): 348
Max. load per shelf (kg / lbs): 40 / 88.2
Permitted total load (kg / lbs): 120 / 264.6
Electrical requirements (380V, 50Hz): 3ph / 12A
No. of wire shelves (standard / max.): 2 / 29
Distance of between shelves (mm / inch): 30 / 1.2
Exhaust air volume flow (approx, L/min): 1910
Dimension (W×D×H) - Exterior (mm / inch): 1500×1145×1912/ 59×45.1×75.3
Dimension (W×D×H) - Interior (mm / inch): 900×800×1000/ 35.4×31.5×39.4
Heating time 2) (min.) from Amb. to 250℃: 55
Dimension (W×D×H) - Net weight (kg / lbs): 390 / 859.8
Temperature fluctuation 1) at 100℃ (±℃ / ℉): 0.3 / 0.54
Temperature variation 2) (±℃ / ℉) - at 100℃: 0.5 / 0.90
Temperature variation 2) (±℃ / ℉) - at 150℃: 0.7 / 1.26
Temperature variation 2) (±℃ / ℉) - at 200℃: 1.1 / 1.98
Temperature variation 2) (±℃ / ℉) - at 250℃: 1.5 / 2.70
Temperature range (℃ / ℉) - intake/damper 100% open: Max. 80 / Max. 176
Temperature range (℃ / ℉) - intake/damper 100% close: Amb.+45 to 250 / Amb. +81 to 482
Recovery time 2) after door was opened (min.) - at 100℃: 6.5
Recovery time 2) after door was opened (min.) - at 150℃: 8
Recovery time 2) after door was opened (min.) - at 200℃: 7
Recovery time 2) after door was opened (min.) - at 250℃: 5.5