Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Alarms
Analog inputs: Hi, Lo, or Hi + Lo evaluated each scan
Digital inputs
34950A/52A digital in maskable pattern match or state change
34950A/52A frequency and totalize: Hi limit only
Monitor channel: Alarm evaluated each reading
Alarm outputs: 4 TTL compatible. Selectable TTL logic Hi or Lo on fail
Latency: 5 ms (typical)
Memory
Type: Volatile
Readings: 500,000 with timestamp, readable during scan
Line power
Power supply: Universal 100 V to 240 V ± 10%
Power line frequency: 50 – 60 Hz ±10% automatically sensed
Power consumption: 150 VA
Environment
Operating environment: Full accuracy for 0 °C to 55 °C
Storage environment - 40°C to 70°C 1
Mechanical
Mainframe dimensions
133 H x 426 W x 341 D mm (5.25" x 16.8" x 14")
Full rack, 3 units high
Mainframe weight: 8.8 kg (19.6 lbs)