KIKUSUI TOS5402 Withstand Voltage And Insulation Resistance Testers (5 kV / 100 mA (500 VA))
Withstanding Voltage Test
AC output section (ACW only)
Output range: 0.05kV to 5.00kV
Output setting accuracy: ±(2% of setting + 20V) (at no load)
Output setting range:
0.000kV to 5.500kV
0.000kV to 1.500kV (When Time Unit is set to Cycle and Phase Shift is set to 0°)
Output resolution: 1V
Max. rated output *1: 500VA (5kV/100mA)
Max. rated voltage: 5kV
Max. rated current: 100mA (when the output voltage is 0.5kV or higher)
Transformer rating: 500VA
Output voltage waveform*2
Waveforms: Sine
Phase shift: The output voltage waveform can start with a phase shifted approx. 90° from the zero-crossing point.
Distortion: 3% or less. (when the output voltage is 0.5kV or higher and no load or a pure resistive load is connected)
Frequency: 50Hz/60Hz
Frequency setting accuracy: ±0.5% (excluding during voltage rise time)
Voltage regulation: ±10% or less (when changing from maximum rated load to no load)
Input voltage variation: ±0.3%(5kV at no load; power supply voltage: 90V to 250V)
Short-circuit current: 200 mA or more (output voltage 1.0 kV or higher)
Output method: PWM switching
Digital voltmeter
Measurement range: AC: 0.000kV to 6.500kV
Display: □. □□□ kV
Accuracy:
V < 500V: ±(1.5% of reading + 10V)
V ≥ 500V: ± 1.5% of reading
Response*1: ACW: True rms/ Mean-value response rms display can be switched.
DCW and IR: Mean-value
Hold feature: Holds the measured voltage value at the end of the test while the PASS/FAIL judgment result is displayed.
Digital ammeter
Measurement range: AC: 0.00mA to 110mA
Display (i = measured current)
i < 10uA: □. □μA
10uA ≦ i < 100uA: □□. □μA
100uA ≦ i < 1mA: □□□. □μA
1mA ≦ i < 10mA: □. □□□mA
10mA ≦ i < 100mA: □□. □□mA
100mA ≦ i: □□□. □mA
Accuracy*1 (i = measured current):
1.00mA ≦ i: ±(1.5% of reading)
i < 1.00mA: ±(1.5% of reading + 20 μA)
Response*2:
ACW: True rms/ Mean-value response rms display can be switched.
Hold feature: The current measurement after a test is finished is held while the pass judgment is displayed.
Insulation Resistance Test
Output voltage: 25Vdc to 1000Vdc, negative polarity
Output setting accuracy: ±(1.0% of setting + 2V)
Output setting range: 0.00kV to 1.02kV
Output setting regulation: 1V
Max. rated load: 1W (-1000Vdc/1mA)
Max. rated current: 1mA
Ripple1000V no load: 2Vp-p or less
Max. rated load: 10Vp-p or less
Voltage regulation: 1 % or less (when changing from maximum rated load to no load)
Short-circuit current: 12mA or less
Discharge function:
Forced discharge after test completion (discharge resistance: approx. 25kΩ)*1
Maximum capacitance: 2μF*2
Discharge time: 0.0 to 100.0 seconds
Discharge function during interlock activation: ON/OFF
Output voltage monitor function: If the output voltage exceeds ±(10% of setting + 10V), the output is turned off, and the protection function is activated.
Digital voltmeter
Measurement function
Measurement range: 0V to -1200V
Display (v = measured voltage)
v < 10V: □ . □ V
10V ≦ v < 100V: □□ . □ V
100V ≦ v < 1000V: □□□ . □ V
1000V ≦ v: □□□□ . □ V
Accuracy: ±(1% of reading + 1V)
Resistance meter
Measurement range / accuracy*1 (i: Measurement current) (R = measured insulation resistance):
40nA ≦ i ≦ 100nA: 200.00MΩ≦R≦10.000GΩ:±(20% of reading)
100nA < i ≦ 200nA:
100.00MΩ≦R < 5.000GΩ:±(10% of reading)
5.000GΩ≦R≦10.000GΩ:±(20% of reading)
200nA < i ≦ 1000nA: 10.000MΩ≦R < 5.000GΩ:±(5% of reading)
1 μA < i ≦ 2mA:
0.001MΩ≦R < 10.000MΩ:±(2% of reading + 5digit)
10.000MΩ≦R < 2.000GΩ:±(2% of reading)
Display (R = measured insulation resistance):
25kΩ ≦ R < 1.000MΩ: □□□kΩ
1.000MΩ ≦ R < 20.0 GΩ:
□ . □□□MΩ
□□ . □□MΩ
□□□ . □MΩ
□ . □□□GΩ
□□ . □□GΩ
20.0GΩ ≦ R: OVER
Hold feature: The resistance measurement after a test is finished is held while the pass judgment is displayed.
Response: Average response (averaged time: Fixed at 100ms)