KIKUSUI TOS5403 Withstand Voltage And Insulation Resistance Testers (6 kV / 10 mA (50 W))
Withstanding Voltage Test
DC output section (DCW only)
Output range: 0.05kV to 6.00kV
Output setting accuracy: ±(2% of setting + 20V) (at no load)
Output setting range: 0.000kV to 6.200kV
Output resolution: 1V
Max. rated output *1: 50W (5kV/10mA)
Max. rated voltage: 6kV
Max. rated current: 10mA
Ripple (TYP)
5kV no load: 50Vp-p
Max. rated load: 100Vp-p
Voltage regulation: 3% or less (when changing from maximum rated load to no load)
Short-circuit current (TYP): 40mA (when generating 6 kV output)
Discharge function:
Forced discharge after test completion (discharge resistance: 125kΩ)*2 Maximum capacitance: 1μF*3
Discharge time: 0.0 to 100.0 seconds
Discharge function during interlock activation: ON/OFF
Start voltage function: The initial voltage at test start can be set to 1% to 99% of the test voltage. Resolution: 1%
Limit voltage function: The upper limit of the test voltage can be set.
ACW: 0.000kV to 5.500kV, DCW: 0.000kV to 6.200kV
Output voltage monitor function: If the output voltage deviates more than ±(350V) form the set value, the output is turned off, and the protection function is activated.
Digital voltmeter
Measurement range: DC: 0.000kV to 6.500kV
Display: □. □□□ kV
Accuracy:
V < 500V: ±(1.5% of reading + 10V)
V ≥ 500V: ± 1.5% of reading
Response*1: Mean-value
Hold feature: Holds the measured voltage value at the end of the test while the PASS/FAIL judgment result is displayed.
Digital ammeter
Measurement range: DC: 0.00mA to 11mA
Display (i = measured current)
i < 10uA: □. □μA
10uA ≦ i < 100uA: □□. □μA
100uA ≦ i < 1mA: □□□. □μA
1mA ≦ i < 10mA: □. □□□mA
10mA ≦ i < 100mA: □□. □□mA
100mA ≦ i: □□□. □mA
Accuracy*1 (i = measured current)
1.00mA ≦ i: ±(1.5% of reading)
i < 1.00mA: ±(1.5% of reading + 20 μA)
Response*2: Mean-value
Hold feature: The current measurement after a test is finished is held while the pass judgment is displayed.
Insulation Resistance Test
Output voltage: 25Vdc to 1000Vdc, negative polarity
Output setting accuracy: ±(1.0% of setting + 2V)
Output setting range: 0.00kV to 1.02kV
Output setting regulation: 1V
Max. rated load: 1W (-1000Vdc/1mA)
Max. rated current: 1mA
Ripple:
1000V no load: 2Vp-p or less
Max. rated load: 10Vp-p or less
Voltage regulation: 1 % or less (when changing from maximum rated load to no load)
Short-circuit current: 12mA or less
Discharge function:
Forced discharge after test completion (discharge resistance: approx. 25kΩ)*1
Maximum capacitance: 2μF*2
Discharge time: 0.0 to 100.0 seconds
Discharge function during interlock activation: ON/OFF
Output voltage monitor function: If the output voltage exceeds ±(10% of setting + 10V), the output is turned off, and the protection function is activated.